AzTrials
AzTrials

Wounds and InjuriesCLCNCT02810080

FiO2 Closed Loop Control Ventilation

A study of Wounds and Injuries and Respiratory Failure, run by Zoll Medical Corporation at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

A randomized, multi-center trial involving critically-ill patients comparing conventional, manual control of inspired oxygen (FiO2) within a target range, with a computer-controlled, closed-loop FiO2 delivery using a continuous ventilator. Read more about this study

Who is it for?
Ages 18 to 65 years
Conditions the study is for
Wounds and Injuries, Respiratory Failure
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
210 participants
Where
1 siteUnited States
Study dates
Dec 2016to Dec 2020
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 65 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • The subject's legally authorized representative will provide signed and dated informed consent.
  • Age 18 - 65, inclusive.
  • Admission to a surgical or neurosurgical intensive care unit following traumatic injury or acute surgical illness
  • Requirement for endotracheal intubation
  • Requirement for mechanical ventilation
  • Patient is currently receiving inspired oxygen concentration (FiO2) >=40%

Who cannot join

  • Age under 18 or over 65
  • Isolated or severe head injury (Glasgow Coma Scale = 6 or less) with expected survival less than 24 hours
  • Brain death
  • Anticipated survival less than 48 hours
  • Pregnant female
  • Patient in whom a pulse oximeter cannot detect a reliable signal due to hypotension, hypothermia, or other injury
Show all 14Show fewer
  • Known carbon monoxide poisoning
  • Uncontrolled diabetic
  • Patient who is unable to maintain SpO2 level of 88% at an FiO2 of 100% due to their medical condition. (NOTE: Being in the trial would impose no additional risk to a patient who presents with this condition however, data from a patient presenting with this pathophysiology would not provide insight into the performance of the controller as it would simply maintain that patient on an FiO2 of 100%.)
  • Patient with chronic hypercarbia.
  • Patient where a physician believes that FiO2 PCLC could be detrimental to the management of the patient.
  • Prisoner
  • SpO2 to SaO2 difference >4%
  • Patient with core temperature <35 C

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Regions Hospital

    Saint Paul, Minnesota, 55101

Other trials in United States

About this study

Official title: FiO2 Closed Loop Control Using the ZOLL 731 Series Ventilator

A randomized, multi-center trial involving critically-ill patients comparing conventional, manual control of inspired oxygen (FiO2) within a target range, with a computer-controlled, closed-loop FiO2 delivery using a continuous ventilator.

Read the detailed description

The trial will use a proportional-integral-derivative (PID) type controller for automatic adjustment of FiO2 to maintain the SpO2 within a target range, based on continuous use of pulse oximetry. Intubated subjects will be randomized to either manual control of FiO2 or closed-loop control of FiO2, during a 12-hour period. The PCLC is designed to respond to acute hypoxemia (SpO2 <88%) within seconds while preventing hypoxemia and minimizing exposure to excessive inspired oxygen levels.

Being tested
Closed Loop Control Ventilation, Manual Control Ventilation
Medical subject headings
Wounds and Injuries, Respiratory Insufficiency

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Closed Loop Control VentilationDevice

    Closed loop control ventilation using ZOLL 731 ventilator with physiologic closed-loop control software

  • Manual Control VentilationDevice

    Manual control ventilation using ZOLL 731 ventilator

Study groups in NCT02810080
Study groupRoleWhat they receive
Closed Loop Control VentilationExperimentalautomatic adjustment of the ZOLL 731 ventilator FiO2 by the closed loop control software
Receives: Device: Closed Loop Control Ventilation
Manual Control VentilationActive comparatormanual adjustment of the ZOLL 731 ventilator FiO2 by the treatment team based on physician order
Receives: Device: Manual Control Ventilation

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Relative Duration With Oxygen Saturation (SpO2) Within Target Range [Time Frame: 12 Hours]

    Relative duration is (time within SpO2 target of 92-96%)/12 hours where higher value demonstrates better ability of FiO2 control method to maintain appropriate oxygen saturation.

    Relative time the patient's SpO2 was maintained within target over the 12 hours on the study device, calculated as total time each patient had SpO2 within the 92 - 96% range divided by 12 hours with arcsin square root transformation

    Measured: 12 hours

  2. Number of Participants With SpO2 Less Than 88% at Anytime While on Study Device

    Number of participants with SpO2 less than 88% at anytime while on study device

    Measured: 24 hours after intervention

Secondary outcomes (1)
  1. Relative Duration of Time With SpO2 92 to 96% When FiO2 is Greater Than 21% and SpO2 is Greater Than or Equal to 92% When Fi02 "Relative Duration With Oxygen Saturation (SpO2) Within Target Range [Time Frame: 12 Hours]

    Relative duration is (time within SpO2 target of either a) 92-96% when FiO2 is greater than 21% or b) greater than or equal to 92% when FiO2 equals 21%)/12 hours where higher value demonstrates better ability of FiO2 control method to maintain appropriate oxygen saturation.

    Relative time the patient's SpO2 was maintained within target over the 12 hours on the study device, calculated as total time each patient had SpO2 within the range divided by 12 hours with arcsin square root transformation

    Measured: 12 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted22 Jun 2016 (estimated)
  2. Study startDec 2016
  3. Primary completionDec 2020
  4. Study completionDec 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
Single blind
Who is masked
Participant
Enrolment
210 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02810080
Sponsor's protocol ID
9703
Acronym
CLC
Status last verified
Apr 2023
First submitted
20 Jun 2016
Last updated
15 Aug 2024
Data monitoring committee
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study of Wounds and Injuries and Respiratory Failure, run by Zoll Medical Corporation at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 15 Aug 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 15 Aug 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Relative Duration With Oxygen Saturation (SpO2) Within Target Range [Time Frame: 12 Hours] (time frame: 12 hours). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 15 Aug 2024.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in