Actis Total Hip System 2 Year Follow-up

A study testing Actis Total Hip System in people with Osteoarthritis and Traumatic Arthritis, run by DePuy Orthopaedics at 9 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Two-year (2 year) information regarding the performance of the commercially available Actis™ Total Hip System in order to obtain and evaluate the clinical outcomes on a series of cementless primary total hip arthroplasty (THA) and hemi-hip arthroplasty procedures using clinical, radiographic and device and procedure… Read more about this study

Who is it for?
Ages 21 years and olderHealthy volunteers can take part
What is being tested?
Actis Total Hip System
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
266 participants
Where
9 sitesUnited States
Study dates
1 Jul 2016to 13 Mar 2024
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Total primary hip replacement is indicated in the following conditions:
  • A severely painful and/or disable joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
  • Avascular necrosis of the femoral head.
  • Acute traumatic fracture of the femoral head or neck.
  • Certain cases of ankylosis.
  • Hemi-hip arthroplasty is indicated in the following conditions:
Show all 22Show fewer
  • Acute fracture of the femoral head or neck that cannot be appropriately reduced and treated with internal fixation.
  • Fracture dislocation of the hip that cannot be appropriately reduced and treated with internal fixation.
  • Avascular necrosis of the femoral head.
  • Non-union of femoral neck fractures.
  • Certain high sub-capital and femoral neck fractures in the elderly.
  • Degenerative arthritis involving only the femoral head in which the acetabulum does not require replacement.
  • Pathology involving only the femoral head in which the acetabulum does not require replacement.
  • Further inclusion criteria for individuals who, in the opinion of the Principal Investigator, are suitable candidates for using the devices specified in this protocol and they are as follows:
  • Individuals who are able to speak, read, and comprehend the informed patient consent (IPC) document and willing and able to provide informed patient consent for participation in the study and have authorized the transfer of his/her information to DePuy Synthes (Joint Reconstruction).
  • Individuals who are willing and able to return for follow-up as specified by the study protocol.
  • Individuals who are a minimum age of 21 years at the time of consent.
  • Individuals who are not bedridden.
  • Individuals who are willing and able to complete the Subject Hip Outcomes questionnaires (i.e., FJS-12, HOOS and Hip Evaluation) as specified by the study protocol.
  • Individuals interested in participating in the parallel and optional Patient Activity Tracking portion of the study are required to link the Patient Activity Tracking device to their own well-functioning smartphone (i.e., iPhone or Android phone) via a downloaded mobile application (i.e., ActiTrak™). If a Subject does not have a smartphone, and would like to participate in the Patient Activity Tracking portion of the study, the Sponsor will be able to provide an Android type tablet to which the required app can be downloaded.
  • Individuals interested in participating in the optional Patient Activity Tracking (ActiTrak™) portion of the study must be willing to wear the Patient Activity Tracking device from the time of consent preoperative (if possible, based on date of consent) and for 3 months postoperative.
  • Individuals interested in participating in the optional Patient Activity Tracking (ActiTrak™) portion of the study must have the ability to return at the 3 month follow-up visit in order to return the Patient Activity Tracking device, and if applicable, the Sponsor's Android tablet back to the study site.

Who cannot join

  • Subjects will be excluded if in the opinion of the Investigator, the individual meets any of the following exclusions:
  • Active local or systemic infection.
  • Loss of musculature, neuromuscular compromise or vascular
  • Poor bone quality, such as osteoporosis, where in the surgeon's opinion, there could be considerable migration of the prosthesis or a significant chance of fracture of the femoral shaft and/or the lack of adequate bone to support the implant(s).
  • Charcot's or Paget's disease.
  • For hemi-hip arthroplasty, any pathological condition of the acetabulum, such as distorted acetabuli with irregularities, protrusion acetabuli (arthrokatadysis), or migrating acetabuli, that would preclude the use of the natural acetabulum as an appropriate articular surface for the hemi-hip prosthesis.
Show all 15Show fewer
  • The Subject is a woman who is pregnant or lactating.
  • Existing contralateral hip that was implanted less than 6 months from the date of surgery at the time of consent into this study.
  • Subjects requiring a simultaneous bilateral hip at the time of consent will be excluded; there can be only one Actis™ study hip. This also means that even if another hip system is used for the contralateral hip during simultaneous THAs, the Subject will be excluded. As noted above, Subjects who have an existing contralateral total hip replacement greater than 6 months postoperatively at the time of consent may be entered into this study if they qualify based upon the eligibility criteria and the approved labeling requirements.
  • Subject had a contralateral amputation.
  • Subject has participated in a clinical investigation with an investigational product (drug or device) in the last three months.
  • Subject is currently involved in any personal injury litigation, medical-legal or worker's compensation claims.
  • Subject, in the opinion of the Investigator, is a drug or alcohol abuser or has a psychological disorder that could affect their ability to complete patient reported questionnaires or be compliant with follow-up requirements.
  • Subject was diagnosed and is taking prescription medications to treat a muscular disorder that limits mobility due to severe stiffness and pain such as fibromyalgia or polymyalgia.
  • Subject has a medical condition with less than 2 years of life expectancy. Note: Diabetes, at present, has not been established as a contraindication. However, because of the increased risk for complications such as infection, slow wound healing, etc., the physician should carefully consider the advisability of hip replacement in the severely diabetic patient.

Trial locations and contacts

9 sites in 1 country.

9 sites in 1 country

United States9 sites
  • Study site

    Austin, Texas

  • Study site

    Boston, Massachusetts

  • Study site

    Charlotte, North Carolina

  • Study site

    Lebanon, New Hampshire

  • Study site

    Louisville, Kentucky, 40241

  • Study site

    New York, New York

  • Study site

    Philadelphia, Pennsylvania

  • Study site

    Renton, Washington

  • Study site

    Rochester, Minnesota

Other trials in United States

About this study

Two-year (2 year) information regarding the performance of the commercially available Actis™ Total Hip System in order to obtain and evaluate the clinical outcomes on a series of cementless primary total hip arthroplasty (THA) and hemi-hip arthroplasty procedures using clinical, radiographic and device and procedure related adverse event assessments.

Read the detailed description

A minimum of 10 study sites.

A prospective, uncontrolled, non-randomized, multicenter study.

A total of 255 Subjects to include a minimum of 225 Primary THA Subjects and a Sub-group with a minimum of 30 hemi-hip arthroplasty Subjects will be prospectively enrolled into the study. All Subjects will be seen for a preoperative clinic visit at the time of consent, and evaluated at the time of surgery, 6 weeks (28-70 days), 3 months (71-132 days), minimum 1 year (304-668 days), and minimum 2 years (669-1,763 days) postoperatively.

Being tested
Actis Total Hip System
Medical subject headings
Osteoarthritis, Arthritis, Rheumatoid, Hip Dislocation, Congenital, Femur Head Necrosis

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Actis Total Hip SystemDevice

    Also called Actis DuoFix Femoral Stem

    Total Hip Replacement and Hemi-hip Replacement

Study groups in NCT02783274
Study groupRoleWhat they receive
Actis Total Hip SystemOtherThe Actis DuoFix Femoral Stem can be used for both a Total and Hemi-hip Replacement
Receives: Device: Actis Total Hip System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Harris Hip Score- THA Subjects Per Protocol at 2 Years

    The Harris Hip Score (HHS) is a standardized tool used to evaluate hip pain, function, and mobility. The HHS is score from 0 to 100, with higher scores indicating better outcomes. A score of less than 70 is considered poor, 70-80 is fair, 80-90 is good, 90-100 is excellent.

    Measured: 2 Years

  2. Harris Hip Score- HHA Subjects Per Protocol at 2 Years

    The Harris Hip Score (HHS) is a standardized tool used to evaluate hip pain, function, and mobility. The HHS is score from 0 to 100, with higher scores indicating better outcomes. A score of less than 70 is considered poor, 70-80 is fair, 80-90 is good, 90-100 is excellent.

    Measured: 2 Years

Secondary outcomes (38)
  1. Harris Hip Score- THA Subjects Per Protocol at 3 Months

    The Harris Hip Score (HHS) is a standardized tool used to evaluate hip pain, function, and mobility. The HHS is score from 0 to 100, with higher scores indicating better outcomes. A score of less than 70 is considered poor, 70-80 is fair, 80-90 is good, 90-100 is excellent.

    Measured: 3 months

  2. Harris Hip Score- HHA Subjects Per Protocol at 3 Months

    The Harris Hip Score (HHS) is a standardized tool used to evaluate hip pain, function, and mobility. The HHS is score from 0 to 100, with higher scores indicating better outcomes. A score of less than 70 is considered poor, 70-80 is fair, 80-90 is good, 90-100 is excellent.

    Measured: 3 months

  3. Harris Hip Score- THA Subjects Per Protocol at 1 Year

    The Harris Hip Score (HHS) is a standardized tool used to evaluate hip pain, function, and mobility. The HHS is score from 0 to 100, with higher scores indicating better outcomes. A score of less than 70 is considered poor, 70-80 is fair, 80-90 is good, 90-100 is excellent.

    Measured: 1 Year

  4. Harris Hip Score- HHA Subjects Per Protocol at 1 Year

    The Harris Hip Score (HHS) is a standardized tool used to evaluate hip pain, function, and mobility. The HHS is score from 0 to 100, with higher scores indicating better outcomes. A score of less than 70 is considered poor, 70-80 is fair, 80-90 is good, 90-100 is excellent.

    Measured: 1 Year

  5. Post-Operative 2-Year Device Survivorship- THA Subjects Safety Population

    Device survivorship at 2 years post-operatively is a secondary endpoint in this study. A revision is defined as a revision for any reason and device survival is defined as the lack of revision. A 2-year Kaplan-Meier survivorship estimate of the Actis Hip System in implanted hips is provided.

    Measured: 2 Years

  6. Post-Operative 2-Year Femoral Stem Survivorship- THA Subjects Safety Population

    A stem revision is defined as any stem revision for any reason and device survival is defined as the lack of revision. A 2-year Kaplan-Meier survivorship estimate of the Actis Hip System in implanted hips is provided.

    Measured: 2 Years

  7. Acetabular Shell Version at 2-years Postoperatively- THA Subjects Per Protocol

    Acetabular cup version is the angle formed between the axis of the implant and the coronal plane of the body. Acetabular version typically ranges between 11.5 and 28.5 degrees.

    Measured: 2 Years

  8. Acetabular Shell Inclination at 2-years Postoperatively- THA Subjects Per Protocol

    Acetabular cup inclination is the angle between the longitudinal axis of the patient and a perpendicular line to the major axis of the cup projection. The target acetabular cup inclination is between 30-50 degrees

    Measured: 2 Years

  9. Number of Participants With >4 Degrees Change in Acetabular Cup Inclination Angle at 2 Years Postoperatively - THA Subjects

    Number of participants with greater than (>) 4 degrees change in acetabular cup inclination angle at 2 years postoperatively was reported.

    Measured: 2 Years

  10. Number of Participants With Acetabular Shell Migration >2mm at 2 Years Postoperatively - THA Subjects Per Protocol

    Number of participants with acetabular shell migration >2mm at 2 years postoperatively were reported. Acetabular migration is the superior migration of the implant within the pelvic bone.

    Measured: 2 Years

  11. Acetabular Radiolucent Lines- THA Subjects Per Protocol

    Radiolucent lines are seen on x-rays and represent an area that is absent of bone. Radiolucent lines seen adjacent to an implant may suggest a lack on bone ingrowth or possible loosening of the implant. Radiolucent lines greater than 1mm wide were considered reportable for this study.

    Measured: 2 Years

  12. Acetabular Osteolysis- THA Subjects Per Protocol

    Osteolysis is seen on x-rays and represent an area where bone has been resorbed by the body in response to small particles of debris that are created as an implant wears over time. Osteolysis greater than 5mm was considered to be reportable for this study.

    Measured: 2 Years

  13. Acetabular Sclerotic Lines- THA Subjects Per Protocol

    Acetabular sclerotic lines are seen on x-rays and are radio-dense areas of bone that have remodeled due to loads close to the implant.

    Measured: 2 Years

  14. Acetabular Porous Coating Integrity- THA Subjects Per Protocol

    Acetabular Cup Porous Coating Integrity is defined as the presence of one or more beads of the porous coating observed away from the component in the soft tissue adjacent to the device.

    Measured: 2 Years

  15. Femoral Stem Position- THA Subjects Per Protocol

    Femoral stem position is categorized as neutral, valgus, or varus. A neutral stem is centered within the femoral canal that runs through the center of the femur's long axis. A valgus femoral stem is positioned slightly towards the medial side of the femoral canal, and a varus femoral stem is positioned slightly towards the lateral side of the femoral canal.

    Measured: 2 Years

  16. Femoral Stem Position- HHA Subjects Per Protocol

    Femoral stem position is categorized as neutral, valgus, or varus. A neutral stem is centered within the femoral canal that runs through the center of the femur's long axis. A valgus femoral stem is positioned slightly towards the medial side of the femoral canal, and a varus femoral stem is positioned slightly towards the lateral side of the femoral canal.

    Measured: 2 Years

  17. Number of Participants With Femoral Stem Tilt - THA Subjects Per Protocol

    Number of participants with femoral stem tilt were reported. Femoral stem tilt is a measurement taken at each post-operative year and compared against baseline images collected at 6-weeks post-operative. The measurement captures changes in the position of the femoral stem over time.

    Measured: 2 Years

  18. Number of Participants With Femoral Stem Tilt- HHA Subjects Per Protocol

    Number of participants with femoral stem tilt were reported. Femoral stem tilt is a measurement taken at each post-operative year and compared against baseline images collected at 6-weeks post-operative. The measurement captures changes in the position of the femoral stem over time.

    Measured: 2 Years

  19. Number of Participants With Femoral Stem Subsidence- THA Subjects Per Protocol

    Femoral stem subsidence is the inferior movement of the stem within the femoral canal. Subsidence greater than 2 mm was considered reportable for this study.

    Measured: 2 Years

  20. Number of Participants With Femoral Stem Subsidence- HHA Subjects Per Protocol

    Femoral stem subsidence is the inferior movement of the stem within the femoral canal. Subsidence greater than 2 mm was considered reportable for this study.

    Measured: 2 Years

  21. Femoral Stem Heterotopic Ossification- THA Subjects Per Protocol

    Heterotopic ossification is the abnormal formation of bone in soft tissues, such as muscles, tendons, or ligaments and can occur around the hip joint, leading to pain, stiffness, and restricted range of motion. It is grade as Class 0 through Class IV, with Class 0 being least impactful and Class IV being most impactful.

    Measured: 2 Years

  22. Femoral Stem Heterotopic Ossification- HHA Subjects Per Protocol

    Heterotopic ossification is the abnormal formation of bone in soft tissues, such as muscles, tendons, or ligaments and can occur around the hip joint, leading to pain, stiffness, and restricted range of motion. It is grade as Class 0 through Class IV, with Class 0 being least impactful and Class IV being most impactful.

    Measured: 2 Years

  23. Femoral Radiolucent Lines- THA Subjects Per Protocol

    Radiolucencies are seen on x-rays and represent an area that is absent of bone. Radiolucent lines seen adjacent to an implant may suggest a lack of bone ingrowth into the implant or possible loosening of the implant.

    Measured: 2 Years

  24. Femoral Radiolucent Lines- HHA Subjects Per Protocol

    Radiolucencies are seen on x-rays and represent an area that is absent of bone. Radiolucent lines seen adjacent to an implant may suggest a lack of bone ingrowth into the implant or possible loosening of the implant.

    Measured: 2 Years

  25. Femoral Osteolysis- THA Subjects Per Protocol

    Osteolysis is seen on x-rays and represent an area where bone has been resorbed by the body in response to small particles of debris that are created as an implant wears over time. Osteolysis greater than 5mm was considered to be reportable for this study.

    Measured: 2 Years

  26. Femoral Osteolysis- HHA Subjects Per Protocol

    Osteolysis is seen on x-rays and represent an area where bone has been resorbed by the body in response to small particles of debris that are created as an implant wears over time. Osteolysis greater than 5mm was considered to be reportable for this study.

    Measured: 2 Years

  27. Femoral Stem Sclerotic Lines- THA Subjects Per Protocol

    Sclerotic lines are seen on x-rays and are radio-dense areas of bone that have remodeled due to loads close to the implant.

    Measured: 2 Years

  28. Femoral Stem Sclerotic Lines- HHA Subjects Per Protocol

    Sclerotic lines are seen on x-rays and are radio-dense areas of bone that have remodeled due to loads close to the implant.

    Measured: 2 Years

  29. Femoral Stem Porous Coating Integrity- THA Subjects Per Protocol

    Acetabular Cup Porous Coating Integrity is defined as the presence of one or more beads of the porous coating observed away from the component in the soft tissue adjacent to the device.

    Measured: 2 Years

  30. Femoral Stem Porous Coating Integrity- HHA Subjects Per Protocol

    Femoral Stem Porous Coating Integrity is defined as the presence of one or more beads of the porous coating observed away from the component in the soft tissue adjacent to the device.

    Measured: 2 Years

  31. Femoral Stem Calcar Resorption- THA Subjects Per Protocol

    Femoral calcar resorption occurs when the bone around the prosthesis is partially shielded from bearing loads and begins to resorb

    Measured: 2 Years

  32. Femoral Stem Calcar Resorption- HHA Subjects Per Protocol

    Femoral calcar resorption occurs when the bone around the prosthesis is partially shielded from bearing loads and begins to resorb

    Measured: 2 Years

  33. Femoral Stem Calcar Fracture- THA Subjects Per Protocol

    Near the top of the femur is a region of bone called the calcar femorale that provides strength to the femur when compressive forces are applied. During preparation of the femur to receive the femoral implant there is the possibility of fractures occurring in this region.

    Measured: 2 Years

  34. Femoral Stem Calcar Fracture- HHA Subjects Per Protocol

    Near the top of the femur is a region of bone called the calcar femorale that provides strength to the femur when compressive forces are applied. During preparation of the femur to receive the femoral implant there is the possibility of fractures occurring in this region.

    Measured: 2 Years

  35. Perceived Leg Length Discrepancy- THA Subjects Per Protocol

    The number of subjects who perceived a Leg Length Discrepancy at Two Years

    Measured: 2 Years

  36. Perceived Leg Length Discrepancy- HHA Subjects Per Protocol

    The number of subjects who perceived a Leg Length Discrepancy at Two Years

    Measured: 2 Years

  37. Number of Participants With True Leg Length Discrepancy- THA Subjects Per Protocol

    True LLD is defined as a difference of 3 cm or greater as reviewed via postoperative radiographs. The measurement is the distance between the superior-most aspects of the left and right femoral heads

    Measured: 2 Years

  38. Number of Participants With True Leg Length Discrepancy- HHA Subjects Per Protocol

    True LLD is defined as a difference of 3 cm or greater as reviewed via postoperative radiographs. The measurement is the distance between the superior-most aspects of the left and right femoral heads

    Measured: 2 Years

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted26 May 2016 (estimated)
  2. Study start1 Jul 2016
  3. Primary completion13 Mar 2024
  4. Study completion13 Mar 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Model description
Single cohort of Total Hips and the protocol also includes a single cohort for Hemi-hips. No comparisons are made to comparator devices, no randomization (open label). gathering data for a Post-market clinical follow-up study.
Masking
None (open label)
Enrolment
266 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02783274
Sponsor's protocol ID
14014
Acronym
Actis14014
Status last verified
Aug 2025
First submitted
23 May 2016
Last updated
15 Aug 2025
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Yes: Johnson \& Johnson Medical Device Companies have an agreement with the Yale Open Data Access (YODA) Project to serve as the independent review panel for evaluation of requests for clinical study reports and participant level data from investigators and physicians for scientific research that will advance medical knowledge and public health. Requests for access to the study data can be submitted through the YODA Project site at http://yoda.yale.edu

Questions about this trial

What is this clinical trial studying?

A study testing Actis Total Hip System in people with Osteoarthritis and Traumatic Arthritis, run by DePuy Orthopaedics at 9 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 22 inclusion and 15 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 15 Aug 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 9 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 22 May 2025). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Harris Hip Score- THA Subjects Per Protocol at 2 Years (time frame: 2 Years). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 15 Aug 2025.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
Actis Total Hip System 2 Year Follow-up (NCT02783274) | AzTrials