Accuracy of LASIK Flaps Created With the LenSx® Femtosecond Laser - A Pilot Study

A study testing Femtosecond Laser System and LASIK surgery in people with Myopia and Hyperopia, run by Alcon Research at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to evaluate the thickness accuracy of Laser-assisted in situ keratomileusis (LASIK) flaps created with the LenSx® Laser. Read more about this study

Who is it for?
Ages 22 years and older
Conditions the study is for
Myopia, Hyperopia
What is being tested?
Femtosecond Laser System and LASIK surgery
Could I receive a placebo?
No, everyone receives an active treatment
How many people
34 participants
Where
1 siteUnited States
Study dates
28 Oct 2015to 9 Apr 2016
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 22 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Diagnosed with myopia or hyperopia and eligible for femtosecond laser initiated LASIK surgery;
  • Best corrected visual acuity (BCVA) correctable to at least 20/25 in each eye;
  • If contact lens wearer, willing to discontinue contact lens wear prior to screening procedures;
  • Willing and able to return for ALL scheduled follow-up examinations;
  • Willing and able to provide written informed consent;
  • Eligible for bilateral LASIK procedures to be done on the same day;
  • Other protocol-specified inclusion criteria may apply.

Who cannot join

  • Previous corneal surgery, corneal lesions that will impede laser treatment, corneal edema, hypotony, glaucoma, existing corneal implant, or keratoconus;
  • Known sensitivity to planned study concomitant medications;
  • Participation in any other ophthalmic drug or device clinical study for 30 days prior to or during the time of participation in this study;
  • Irregular astigmatism, based on the Investigator's judgement;
  • Pregnant, lactating or planning to become pregnant during the course of the study;
  • Other protocol-specified exclusion criteria may apply.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Contact Alcon Call Center for Trial Locations

    Fort Worth, Texas, 76134

Other trials in United States

About this study

The purpose of this study is to evaluate the thickness accuracy of Laser-assisted in situ keratomileusis (LASIK) flaps created with the LenSx® Laser.

Conditions
Myopia, Hyperopia
Being tested
Femtosecond Laser System, LASIK surgery
Medical subject headings
Myopia, Hyperopia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Femtosecond Laser SystemDevice

    Also called LenSx®

    Used for the creation of a corneal flap in patients undergoing LASIK surgery

  • LASIK surgeryProcedure or surgery

Study groups in NCT02575911
Study groupRoleWhat they receive
LenSxExperimentalLASIK surgery in both eyes using LenSx® Femtosecond Laser System
Receives: Device: Femtosecond Laser System; Procedure: LASIK surgery

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Flap Thickness Accuracy Within the Central Zone at Month 3 Postoperative

    Flap thickness was assessed by optical coherence tomography (OCT) and averaged from 3 separate scans. Flap thickness accuracy was defined as the mean difference between the achieved and desired (desired subtracted from achieved) flap thickness. A positive number represents a postoperative flap thickness that is thicker than the expected flap thickness and vice versa for a negative number. Each eye contributed separately to the analysis.

    Measured: Month 3 postoperative

Secondary outcomes (9)
  1. Flap Thickness Accuracy Within the Central Zone at Month 1 Postoperative

    Flap thickness was assessed by OCT and averaged from 3 separate scans. Flap thickness accuracy was defined as the mean difference between the achieved and desired (desired subtracted from achieved) flap thickness. A positive number represents a postoperative flap thickness that is thicker than the expected flap thickness and vice versa for a negative number. Each eye contributed separately to the analysis.

    Measured: Month 1 postoperative

  2. Flap Thickness Precision Within the Central Zone at Month 1 and Month 3 Postoperative

    Flap thickness was assessed by OCT and averaged from 3 separate scans. Flap thickness precision (variability of the achieved flap thickness) was defined as the standard deviation of the flap thickness measurement. Each eye contributed separately to the analysis.

    Measured: Month 1, Month 3 postoperative

  3. Ease of Flap Dissection at Day 0, Operative Day

    Ease of flap dissection (ease of lifting the flap during surgery) was assessed on a scale from 0 to 5 where 0 = Unable to lift flap and 5 = Able to lift flap without any resistance using blunt instrument. Each eye contributed separately to the analysis.

    Measured: Day 0, operative day

  4. Stromal Bed Quality at Day 0, Operative Day

    Stromal bed quality was assessed by the investigator on a roughness scale from 0 to 5 where 0 = Very rough surface and 5 = Very smooth surface. Each eye contributed separately to the analysis.

    Measured: Day 0, operative day

  5. Opaque Bubble Layer (OBL) at Day 0, Operative Day

    Opaque bubble layer was assessed by the investigator during the surgery on a scale from 0 to 5 where 0 = No OBL and 5 = 100% OBL in the stromal bed area. Each eye contributed separately to the analysis.

    Measured: Day 0, operative day

  6. Uncorrected Distance Visual Acuity (UCDVA) at Month 1 and Month 3 Postoperative

    UCDVA (measurement of uncorrected (without spectacles or other visual corrective devices) distance visual acuity) was assessed at a distance of 6 meters or 20 feet using a standard Snellen eye chart, both monocularly and binocularly, and reported categorically as a percentage of eyes analyzed.

    Measured: Month 1, Month 3 postoperative

  7. Best Corrected Distance Visual Acuity (BCDVA) by Visit

    BCDVA (measurement with the participant's best spectacle correction) was assessed at a distance of 6 meters or 20 feet using a standard Snellen eye chart, both monocularly and binocularly, and reported categorically as a percentage of eyes analyzed.

    Measured: Baseline, Month 1, Month 3 postoperative

  8. Manifest Refraction Spherical Equivalent (MRSE)

    The participant was manually refracted to his/her best correction using a phoropter and standard Snellen eye charts. Each eye contributed separately to the analysis.

    Measured: Baseline, Month 1, Month 3 postoperative

  9. Prediction Error Between Target Versus Achieved Refraction at One and Three Months Post-op

    Prediction error was summarized as a percentage of eyes within 0.5 D and 1.0 D of target at one and three months postoperative

    Measured: Month 1, Month 3 postoperative

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted15 Oct 2015 (estimated)
  2. Study start28 Oct 2015
  3. Primary completion9 Apr 2016
  4. Study completion9 Apr 2016

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
34 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02575911
Sponsor's protocol ID
CTJ121-P001
Status last verified
Apr 2017
First submitted
13 Oct 2015
Last updated
2 Jul 2018
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes

Questions about this trial

What is this clinical trial studying?

A study testing Femtosecond Laser System and LASIK surgery in people with Myopia and Hyperopia, run by Alcon Research at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 2 Jul 2018. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 24 May 2017). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Flap Thickness Accuracy Within the Central Zone at Month 3 Postoperative (time frame: Month 3 postoperative). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 2 Jul 2018.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in