Oximetry, PulseNCT02571686

Accuracy of Pulse Oximeters With Profound Hypoxia

An observational study of Oximetry, Pulse, run by Mortara Instrument at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The aim of this project is to test the accuracy of pulse oximeters during mild, moderate and severe hypoxia. This is done by comparing the reading of the pulse oximeter during brief, steady state hypoxia with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a laboratory… Read more about this study

Who is it for?
Ages 18 to 50 yearsHealthy volunteers can take part
Conditions the study is for
Oximetry, Pulse
What is being tested?
Surveyor S4
Type of treatment
Medical device
How many people
10 participants
Where
1 siteUnited States
Study dates
Oct 2015to Oct 2015
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 50 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Study population: 10-12 healthy volunteers

Who may be able to join

  • The subject is male or female, aged ≥18 and <50.
  • The subject is in good general health with no evidence of any medical problems.
  • The subject is fluent in both written and spoken English.
  • The subject has provided informed consent and is willing to comply with the study procedures.

Who cannot join

  • The subject is obese (BMI>30).
  • The subject has a known history of heart disease, lung disease, kidney or liver disease.
  • Diagnosis of asthma, sleep apnea, or use of CPAP(Continuous positive airway pressure).
  • Subject has diabetes.
  • Subject has a clotting disorder.
  • The subject a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.
Show all 16Show fewer
  • The subject has any other serious systemic illness.
  • The subject is a current smoker.
  • Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly.
  • The subject has a history of fainting or vasovagal response.
  • The subject has a history of sensitivity to local anesthesia.
  • The subject has a diagnosis of Raynaud's disease.
  • The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
  • The subject is pregnant, lactating or trying to get pregnant.
  • The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.
  • The subject has any other condition, which in the opinion of the investigators' would make them unsuitable for the study -

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Department of Anesthesia Clinical Studies Facility, room A-65 Parnassus Campus.

    San Francisco, California, 94143

Other trials in United States

About this study

The aim of this project is to test the accuracy of pulse oximeters during mild, moderate and severe hypoxia. This is done by comparing the reading of the pulse oximeter during brief, steady state hypoxia with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a laboratory hemoximeter). The data obtained is submitted by pulse oximeter manufacturers to the FDA for device approval.

Read the detailed description

Overall design:

Studies normally involve 6-12 normal adult volunteer paid subjects, with 20-25 1-2 ml arterial blood samples from each subject obtained at different steady-state levels of hypoxia from 70-100%. Blood samples obtained from an arterial line are measured in a hemoximeter to determine true saturation value. Some sponsors may not require any blood sampling. Some sponsors may require additional blood samples, up to 35 samples may be drawn.

Readings from the test pulse oximeters are recorded and compared to these "gold-standard" blood values. The pulse oximeter probes may be located at a variety of sites on the subjects, including fingers, toes, ears, forehead, scalp and bridge of the nose. These detector-probes are all non-invasive. The level of hypoxia is measured and controlled by the investigator. A computer program that displays a prediction, breath by breath, the arterial oxygen saturation, (SaO2) of the study subject. SaO2 is computed from end-expired Po2 and Pco2 as determined by mass spectrometer gas analysis. This information permits the inspired gas mixture of air, plus CO2 and nitrogen, to be adjusted by an operator watching the value computed after each expiration on an analog meter. This computer-estimated saturation is adjusted by the operator to one of 6 levels of predicted saturation, and is held stable for about 30 seconds at each level. Two or three "runs" are conducted per subject. Each "run" lasts 8-12 min and 4 to 5 plateaus are tested per run. The manufacturer may choose the target values. Plateaus are typically sought at 92%, 86%, 80%, 74%, 68% and 62%. Other manufacturers have asked for an equal number of data points but with all points between 70% and 100%. Some sponsors may also request additional variables to be measured during testing including high or low Carbon Dioxide (CO2), where subjects will be asked to hyperventilate; low perfusion, where subjects will be asked to lay flat, or with their head up or down; testing pulse oximeters during motion, with the subject's hand fixed to a motion machine; or measure eye-tracking during hypoxia. The subject's will be informed by the study staff and in the consent form if any of these additional procedures apply.

.

Conditions
Oximetry, Pulse
Being tested
Surveyor S4

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Surveyor S4Device

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Accuracy of Pulse Oximeter

    Measured: 2 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted8 Oct 2015 (estimated)
  2. Study startOct 2015
  3. Primary completionOct 2015
  4. Study completionOct 2015

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Time perspective
Prospective
Enrolment
10 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02571686
Sponsor's protocol ID
139253
Status last verified
Dec 2015
First submitted
6 Oct 2015
Last updated
24 Dec 2015
Results posted
No
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

An observational study of Oximetry, Pulse, run by Mortara Instrument at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 24 Dec 2015. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Accuracy of Pulse Oximeter (time frame: 2 months). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Dec 2015.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in