New Generation Enlite Accuracy Study

A study testing Glucose sensor (Enlite) in people with Diabetes, run by Medtronic MiniMed, Inc..

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The trial will investigate the accuracy and performance of a new sensor for continuous glucose monitoring (CGM), the (new generation) Enlite, CE-marked in 2013, in conjunction with the new Medtronic 640G system components. To evaluate the accuracy and performance of the new generation Enlite sensor. Read more about this study

Who is it for?
Ages 18 to 74 years
Conditions the study is for
Diabetes
What is being tested?
Glucose sensor (Enlite)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
24 participants
Where
No sites listed yet
Study dates
Apr 2015to Sep 2015
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 74 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subject has signed the informed consent form prior to screening
  • Age 18-74 years (both inclusive)
  • Subject has been diagnosed with T1DM
  • Subject is current insulin pump user for at least 3 months
  • HbA1c 9.5%
  • With the exception of T1DM, the subject is considered generally healthy upon completion of medical history, physical examination, vital signs, ECG and biochemical investigations as judged by the Investigator

Who cannot join

  • Female of childbearing potential who is pregnant, breast-feeding or intend to become pregnant or are not using adequate contraceptive methods (adequate contraceptive measures include sterilization, hormonal intrauterine devices, oral contraceptives, sexual abstinence, vasectomized partner)
  • Subject has any condition that, in the opinion of the Investigator or qualified Investigational Centre staff, may preclude him/her from participating in the study and completing study related procedures.
  • Subject has impaired vision or hearing problems that could compromise the handling of the device as determined by Investigator or qualified Investigational Centre staff
  • Subject is unable to tolerate tape adhesive in the area of sensor placement
  • Subject has any unresolved adverse skin condition in the area of sensor placement (e.g. psoriasis, rash, Staphylococcus infection)
  • Subject has travel plans which would make it difficult for the subject to attend on-site study visits as scheduled
Show all 10Show fewer
  • Subject is actively participating in an investigational study (drug or device) wherein he/she has received treatment from an investigational study drug or investigational study device in the last 30 days
  • Subject known to be positive for Hepatitis B surface antigen (HBsAg) or Hepatitis C antibodies (or diagnosed with active hepatitis according to local practice).
  • Positive result to the screening test for HIV-1 antibodies, HIV-2 antibodies or HIV-1 antigen according to locally used diagnostic testing
  • Significant history of alcoholism or drug/chemical abuse as per investigator's judgment, current regular alcohol consumption of more than 1 unit per day for women and more than 2 units per day for men [1 unit of alcohol equals 1 beer (330 mL), 1 glass of wine (120 mL), or 1 glass of spirits (40 mL)], or a positive result in the urine drug/alcohol screen at the screening visit

About this study

Official title: Accuracy and Performance Evaluation of the Medtronic New Generation Enlite Glucose Sensor in Conjunction With the Medtronic 640G System Components - in Clinic and at Home

The trial will investigate the accuracy and performance of a new sensor for continuous glucose monitoring (CGM), the (new generation) Enlite, CE-marked in 2013, in conjunction with the new Medtronic 640G system components.

To evaluate the accuracy and performance of the new generation Enlite sensor. This is a postmarket, interventional, open-label, non-randomized, single center clinical research trial.

The trial will adopt a prospective single-sample design using the Yellow Spring Instruments (YSI) plasma glucose reference sampling procedure in the clinic and self-monitoring of blood glucose (SMBG) plasma glucose reference sampling procedure at home.

Read the detailed description

The trial will investigate the accuracy and performance (i.e. the life-time) of a new sensor for CGM, the (new generation) Enlite glucose sensor, CE-marked in 2013, in conjunction with the new Medtronic 640G system components. Accuracy and life-time of the sensor are important parameters contributing to the safety of patients and patient acceptance of this still fairly new technology.

The current trial investigates the accuracy of the new sensor in conjunction with the new 640G system components both in the clinic and at home. This enables us to assess the sensor accuracy in a well-controlled environment using a highly accurate laboratory glucose analyzer as reference as well as sensor accuracy during real-life conditions using a blood glucose meter as reference analyzer.

The data gathered in this trial will help to understand the effect of technological advancements made and can guide future development of glucose sensor for continuous glucose monitoring.

Conditions
Diabetes
Being tested
Glucose sensor (Enlite)
Keywords
Diabetes, Glucose sensor, Sensor accuracy and performance, CGM
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Glucose sensor (Enlite)Device

    This is a postmarket, interventional, open-label, non-randomized, single center clinical research trial. The trial will adopt a prospective single-sample design using the Yellow Spring Instruments (YSI) plasma glucose reference sampling procedure in the clinic and selfmonitoring of blood glucose (SMBG) plasma glucose reference sampling procedure at home. The trial will consist of 6 visits to the investigational site: a visit for informed consent procedures (Visit 0), a screening visit (Visit 1), 3 study visits (Visit 2-4) and a final examination (Visit 5).

Study groups in NCT02423798
Study groupRoleWhat they receive
Open-label, single-sample designOtherThe trial has an open label design with a glucose sensor intervention. Only 1 sensor system is used in the trial (no comparator). As all subjects will receive identical devices and undergo the same experimental procedures, no randomization will be performed. Furthermore, neither subjects nor clinical staff will be blinded to the sensor readings, as knowing the sensor glucose readings will not affect the accuracy endpoint.
Receives: Device: Glucose sensor (Enlite)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Sensor Accuracy

    Sensor values were compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing day (day 3). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100).

    Measured: 4 months

  2. Consensus Error Grid Analysis of Paired Sensor and Reference Plasma Glucose Values

    All analysis performed using the Consensus Error Grid (or Parkes error grid) comparing the paired sensor and YSI reference glucose values.

    Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action." Zone B as "altered clinical action with little or no effect on clinical outcome.". Ideal situation is 100% in Zone A + B.

    Measured: 4 months

  3. Sensor Survival

    sensor survival in hours

    Measured: 4 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted22 Apr 2015 (estimated)
  2. Study startApr 2015
  3. Primary completionSep 2015
  4. Study completionSep 2015

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
24 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02423798
Sponsor's protocol ID
CEP 296
Status last verified
Jan 2018
First submitted
23 Mar 2015
Last updated
17 Jan 2018
Data monitoring committee
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Glucose sensor (Enlite) in people with Diabetes, run by Medtronic MiniMed, Inc..

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 17 Jan 2018. Status can change between updates, so confirm with the study contact before making plans.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 25 Sep 2017). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Sensor Accuracy (time frame: 4 months). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Jan 2018; AzTrials last refreshed its copy on 29 Oct 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in