Type 1 DiabetesE3NCT02246582
Adult Accuracy Study of the Enlite 3 Glucose Sensor
A study testing Enlite 3, Guardian Mobile App and 640G Insulin Pump in people with Type 1 Diabetes and Type 2 Diabetes, run by Medtronic MiniMed, Inc. at 6 sites in United States.
Trial summary
The purpose of this study is to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes. Read more about this study
- Who is it for?
- Ages 14 to 75 years
- Conditions the study is for
- Type 1 Diabetes, Type 2 Diabetes
- What is being tested?
- Enlite 3, Guardian Mobile App and 640G Insulin Pump
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 118 participants
- Where
- 6 sitesUnited States
- Study dates
- Sep 2014to Aug 2015
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 14 to 75 years
- All sexes
- No healthy volunteers
- Child, Adult, Older adult
Who may be able to join
- Subject is 14-75 years of age at time of screening
- A clinical diagnosis of type 1 or type 2 diabetes for a minimum of 12 months duration as determined via medical record or source documentation by an individual qualified to make a medical diagnosis
- Adequate venous access as assessed by investigator or appropriate staff
- Subjects participating in the high and low glucose challenges must have an established insulin carbohydrate ratio(s) and insulin sensitivity ratio. (The term "established" refers to a ratio that has been previously defined and tested prior to screening visit). Subjects without established ratios may participate under observation only.
Who cannot join
- Subject will not tolerate tape adhesive in the area of Enlite 3 Sensor placement as assessed by qualified individual
- Subject has any unresolved adverse skin condition in the area of the Enlite 3 Sensor or device placement (e.g. psoriasis, rash, Staphylococcus infection)
- Subject is actively participating in an investigational study (drug or device) in the last 2 weeks
- Subject is female and has a positive pregnancy screening test
- Females of child bearing age and who are sexually active should be excluded if they are not using a form of contraception deemed reliable by the investigator
- Subject is female and plans to become pregnant during the course of the study
Show all 14Show fewer
- Subject has had a hypoglycemic seizure within the past 6 months
- Subject has had hypoglycemia resulting in loss of consciousness within the past 6 months prior to screening visit
- Subject has had an episode of diabetic ketoacidosis (DKA) within the past 6 months prior to screening visit
- Subject has a history of a seizure disorder
- Subject has central nervous system or cardiac disorder resulting in syncope
- Subject has a history of myocardial infarction, unstable angina, coronary artery bypass surgery, coronary artery stenting, transient ischemic attack (TIA), cerebrovascular accident (CVA), angina, congestive heart failure, ventricular rhythm disturbances or thromboembolic disease
- Subject has a hematocrit (Hct) lower than the normal reference range
- Subject has a history of adrenal insufficiency
Trial locations and contacts
6 sites in 1 country.
6 sites in 1 country
United States6 sites
Atlanta Diabetes Associates
Atlanta, Georgia, 30318
Barbara Davis Center for Childhood Diabetes
Aurora, Colorado, 80045
Barbara Davis Center
Aurora, Colorado, 80045
AMCR Institute
Escondido, California, 92026
Rainier Clinical Research Center
Renton, Washington, 98057
Diablo Clinical Research
Walnut Creek, California, 94598
About this study
Official title: A Performance Evaluation of the Enlite® 3 Sensor to Support a Full 168 Hours (7 Days) of Use
The purpose of this study is to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes.
Read the detailed description
This study is a multi-center, randomized, in-clinic trial that aims to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes for at least one year.
Subjects are randomized into either Group A or Group B. The groups assigned will determine when the subject will be participating in the in-clinic YSI frequent sample testing (FST). For example on Day 1: Group A begins FST 30 minutes after Enlite 3 Sensor Insertion. Group B begins FST 14 hours after Enlite 3 Sensor Insertion.
- Conditions
- Type 1 Diabetes, Type 2 Diabetes
- Being tested
- Enlite 3, Guardian Mobile App, 640G Insulin Pump
- Keywords
- Type 1 Diabetes, Type 2 Diabetes, Enlite 3
- Medical subject headings
- Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Enlite 3Device
Use of Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen \& arm used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years who have had a diagnosis of type 1 or type 2 diabetes for at least one year.
Guardian Mobile AppDevice
640G Insulin PumpDevice
| Study group | Role | What they receive |
|---|---|---|
| Group A | Other | Subjects underwent FST at 30 mins, 50 hrs and 146 hrs from Enlite 3 Sensors connected to GST3C, GST4C Transmitter, and GSR Receives: Device: Enlite 3; Device: Guardian Mobile App; Device: 640G Insulin Pump |
| Group B | Other | Subjects underwent FST at 14 hrs, 62 hrs and 158 hrs from Enlite 3 Sensors connected to GST3C, GST4C Transmitter, and GSR Receives: Device: Enlite 3; Device: Guardian Mobile App; Device: 640G Insulin Pump |
What is being measured
The primary outcome is the main question the study is designed to answer.
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD)
Enlite 3 Sensor accuracy using two real time devices: 1) 640G Pump and 2) Guardian Mobile with the minimum calibration requirements (every 12 hours after the second calibration) will be evaluated. Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Note that results from multiple testing days will be pooled together for reporting purpose.
Measured: 7 Days
Secondary outcomes (2)
Retrospective Re-Analysis (MARD With 1 Additional Calibration)
Retrospective re-analysis to simulate 640G Pump and Guardian Mobile 1-minute raw data collected by GST3C Transmitters and GST4C Transmitter: Enlite 3 Sensor accuracy with 3-4 calibrations throughout the day (derived from re-analysis of Enlite 3 Sensor data using actual fingerstick values). Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Note that results from multiple testing days will be pooled together for reporting purpose.
Measured: 7 Days
Retrospective Analysis (MARD for the GSR With Minimum and 1 Additional Calibration)
Retrospective analysis using one GSR: minimum and 3-4 calibrations will be evaluated. Enlite 3 Sensor values will be compared to YSI plasma glucose values during YSI frequent sample testing. Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Note that results from multiple testing days will be pooled together for reporting purpose.
Measured: 7 Days
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted23 Sep 2014 (estimated)
- Study startSep 2014
- Primary completionAug 2015
- Study completionAug 2015
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Type 1 DiabetesStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Enrolment
- 118 participants (actual)
Sponsor and investigators
- Lead sponsor
- Medtronic MiniMed, Inc.
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Ronald Brazg, M.D, Principal investigator, Rainier Clinical Research CenterMark Christiansen, M.D, Principal investigator, Diablo Clinical ResearchTimothy Bailey, M.D, Principal investigator, AMCR InstituteSatish Garg, M.D, Principal investigator, Barbara Davis CenterRobert Slover, M.D, Principal investigator, Barbara Davis CenterBruce Bode, M.D, Principal investigator, Atlanta Diabetes Associates
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02246582
- Sponsor's protocol ID
- CEP292
- Acronym
- E3
- Status last verified
- Oct 2016
- First submitted
- 11 Sep 2014
- Last updated
- 2 Nov 2016
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- Yes
Publications
Christiansen MP, Garg SK, Brazg R, Bode BW, Bailey TS, Slover RH, Sullivan A, Huang S, Shin J, Lee SW, Kaufman FR. Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor. Diabetes Technol Ther. 2017 Aug;19(8):446-456. doi: 10.1089/dia.2017.0087. Epub 2017 Jul 12.
PubMed 28700272
Study at a glance
From the registry record, last updated by the sponsor on 2 Nov 2016. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- A Performance Evaluation of the Enlite® 3 Sensor to Support a Full 168 Hours (7 Days) of Use
- ClinicalTrials.gov ID
- NCT02246582
- Acronym / sponsor ID
- E3 / CEP292
- Conditions
- Type 1 Diabetes, Type 2 Diabetes
- Keywords
- Type 1 Diabetes, Type 2 Diabetes, Enlite 3
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 118 participants (actual)
- Lead sponsor
- Medtronic MiniMed, Inc.Industry (pharma, biotech, devices)
- Study start
- Sep 2014
- Primary completion
- Aug 2015 (actual)
- Study completion
- Aug 2015 (actual)
- First posted
- 23 Sep 2014 (estimated)
- Results
- Posted 2 Nov 2016
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of this study is to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes.
Detailed description
This study is a multi-center, randomized, in-clinic trial that aims to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes for at least one year.
Subjects are randomized into either Group A or Group B. The groups assigned will determine when the subject will be participating in the in-clinic YSI frequent sample testing (FST). For example on Day 1: Group A begins FST 30 minutes after Enlite 3 Sensor Insertion. Group B begins FST 14 hours after Enlite 3 Sensor Insertion.
Study design
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Group A | Other | Device: Enlite 3, Device: Guardian Mobile App, Device: 640G Insulin PumpSubjects underwent FST at 30 mins, 50 hrs and 146 hrs from Enlite 3 Sensors connected to GST3C, GST4C Transmitter, and GSR |
| Group B | Other | Device: Enlite 3, Device: Guardian Mobile App, Device: 640G Insulin PumpSubjects underwent FST at 14 hrs, 62 hrs and 158 hrs from Enlite 3 Sensors connected to GST3C, GST4C Transmitter, and GSR |
Who may qualify
- Ages
- 14 to 75 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Child, Adult, Older adult
Key inclusion criteria (4 listed)
- Subject is 14-75 years of age at time of screening
- A clinical diagnosis of type 1 or type 2 diabetes for a minimum of 12 months duration as determined via medical record or source documentation by an individual qualified to make a medical diagnosis
- Adequate venous access as assessed by investigator or appropriate staff
- Subjects participating in the high and low glucose challenges must have an established insulin carbohydrate ratio(s) and insulin sensitivity ratio. (The term "established" refers to a ratio that has been previously defined and tested prior to screening visit). Subjects without established ratios may participate under observation only.
Key exclusion criteria (14 listed)
- Subject will not tolerate tape adhesive in the area of Enlite 3 Sensor placement as assessed by qualified individual
- Subject has any unresolved adverse skin condition in the area of the Enlite 3 Sensor or device placement (e.g. psoriasis, rash, Staphylococcus infection)
- Subject is actively participating in an investigational study (drug or device) in the last 2 weeks
- Subject is female and has a positive pregnancy screening test
- Females of child bearing age and who are sexually active should be excluded if they are not using a form of contraception deemed reliable by the investigator
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Enlite 3Use of Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen \& arm used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years who have had a diagnosis of type 1 or type 2 diabetes for at least one year. | Device | Group A, Group B |
| Guardian Mobile App | Device | Group A, Group B |
| 640G Insulin Pump | Device | Group A, Group B |
Primary outcome
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD)
Time frame: 7 Days
Enlite 3 Sensor accuracy using two real time devices: 1) 640G Pump and 2) Guardian Mobile with the minimum calibration requirements (every 12 hours after the second calibration) will be evaluated. Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Note that results from multiple testing days will be pooled together for reporting purpose.
2 secondary outcomes
Retrospective Re-Analysis (MARD With 1 Additional Calibration)
Time frame: 7 Days
Retrospective Analysis (MARD for the GSR With Minimum and 1 Additional Calibration)
Time frame: 7 Days
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 2 Nov 2016.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Derived from this study: PubMed 28700272Christiansen MP, Garg SK, Brazg R, Bode BW, Bailey TS, Slover RH, Sullivan A, Huang S, Shin J, Lee SW, Kaufman FR. Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor. Diabetes Technol Ther. 2017 Aug;19(8):446-456. doi: 10.1089/dia.2017.0087. Epub 2017 Jul 12.
Related trials for Type 1 Diabetes
Not yet recruitingPhase 3
A Study to Evaluate the Effectiveness and Safety of the Drug AZD6234 for Managing Weight in Adults Who Are Obese or Overweight. These Participants May or May Not Also Have Type 2 Diabetes
Obesity, Overweight
RecruitingPhase 1
Pharmacokinetics, Safety, and Immunogenicity Comparison of Bmab1700 and Opdivo® as Adjuvant Monotherapy in Participants With Melanoma
Melanoma
Active, not recruitingPhase 2
A Study of Botensilimab and Balstilimab for the Treatment of Colorectal Cancer
Metastatic Colorectal Cancer
Not yet recruiting
Inter-Arm Differences in Continuous Glucose Monitor Readings in Adults With Diabetes
Continuous Glucose Monitoring, Type 1 Diabetes
RecruitingObservational study
Estimated Risk of Preeclampsia in the First Trimester and Its Association With Subclinical Atherosclerosis and Cardiovascular Risk in Women With Type 1 Diabetes
Type 1 Diabetes, Preeclampsia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Ronald Brazg, M.D principal investigator
Rainier Clinical Research Center
Mark Christiansen, M.D principal investigator
Diablo Clinical Research
Timothy Bailey, M.D principal investigator
AMCR Institute
6 sites in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Enlite 3, Guardian Mobile App and 640G Insulin Pump in people with Type 1 Diabetes and Type 2 Diabetes, run by Medtronic MiniMed, Inc. at 6 sites in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 2 Nov 2016. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 6 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 2 Nov 2016). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD) (time frame: 7 Days). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT02246582, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 2 Nov 2016.