RF Surgical Sponge-Detecting System on the Function of Pacemakers and Implantable Cardioverter Defibrillators

A study testing RF Assure Scanning in people with Cardiomyopathy and Cardiac Arrhythmia, run by University of California, San Diego at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

If a surgical sponge is mistakenly left inside a patient's body after a surgical procedure, it can cause a serious infection. To prevent this from happening, a new device has been developed that uses radiofrequency (RF) signals to detect the presence of surgical sponges inside the body. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Cardiomyopathy, Cardiac Arrhythmia, Heart Failure
What is being tested?
RF Assure Scanning
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
50 participants
Where
1 siteUnited States
Study dates
Mar 2014to May 2015
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients must be at least 18 years of age
  • Patients must be willing and able to provide consent
  • Patients must be eligible for pacemaker/ICD removal or temporary pacemaker

Who cannot join

  • Pregnancy
  • Inability or unwillingness to comply with the protocol
  • Medical condition that would limit study participation

Trial locations and contacts

1 site in 1 country.

United States1 site
  • UCSD Sulpizio Cardiovascular Center

    La Jolla, California, 92093

Other trials in United States

About this study

Official title: Compatibility of Radiofrequency Detection Technology With Cardiac Internal Electronic Devices (RF Assure Study)

If a surgical sponge is mistakenly left inside a patient's body after a surgical procedure, it can cause a serious infection. To prevent this from happening, a new device has been developed that uses radiofrequency (RF) signals to detect the presence of surgical sponges inside the body. The device is now being used routinely to make sure that no sponges are left inside a patient at the end of an operation. However, the RF device has not been implemented in procedures for patients with cardiac implantable electronic devices (CIEDs). While the device is FDA approved for use, there is a theoretical concern that the radiofrequency signals used to detect the sponges will change the settings on the pacemaker or the defibrillator. Changing the settings on a pacemaker might make it pace the heart too quickly or too slowly, while changing the settings on a defibrillator might cause unnecessary shocks or prevent it from shocking the heart if the patient were to have cardiac arrest.

The purpose of this study is to test whether the radiofrequency device used to detect sponges can cause a clinically significant change to the settings on pacemakers and defibrillators. To minimize potential risk, the device will be tested only on patients who are having the pacemaker or defibrillator removed or replaced as part of their regular medical care, either because it is infected or because the battery has worn out. Before the pacemaker or defibrillator is removed, the settings will be carefully and completely recorded and the radiofrequency device will be used to scan the body for sponges as it would be done during normal operation.

After the pacemaker or defibrillator is taken out, the settings will again be recorded and compared to the settings before the scan. In a standard device removal procedure, no clinically significant change in CIED settings would be expected. If a new pacemaker or defibrillator is implanted in the patient, it will not be exposed to the detection device at all. We will also test whether the RF device has any effect on temporary pacemakers that patients may receive after open heart surgery. We plan to perform testing in a total of 50 patients, 40 with permanent pacemakers or defibrillators and 10 with temporary pacemakers.

Read the detailed description

Retained surgical items (e.g., sponges, needles, and instruments) are among the most frequently reported medical errors and occur in an estimated 1 in 5000 operations. A retained surgical item frequently causes a serious negative outcome, including re-operation (70-80% of cases), readmission or prolonged hospital stay (30-60% of cases), sepsis (43% of cases), or death (2% of cases). Retained surgical items are classed as "never events" by the Centers for Medicare and Medicaid Services (CMS), and CMS does not reimburse for additional medical care related to the treatment for such a complication.

Sponges are by far the most common retained items and are estimated to account for 50-70% of cases. The standard method to ensure that no sponge is left inside the body is manual counting. Manual counting, however, carries a sensitivity and specificity of only about 77% and 99%, respectively. In one study 62% of retained items were detected after the count was reported as normal. If the manual count suggests that a sponge has been retained, the standard method to detect it is by taking a radiograph. Radiographs, however, are time-consuming and in one large retrospective study failed to detect retained items in 33% of cases. The standard methods of detecting retained surgical sponges are clearly suboptimal.

Radiofrequency (RF) technology has recently been employed to improve the detection of retained surgical sponges. A radiofrequency chip is sewn into the fabric of the sponge, and a circular wand that emits a radiofrequency signal is passed over the patient. The wand also serves as an antenna that detects a return signal from the chip in the sponge. If a chip is detected it triggers an audio and a visual alarm on a console attached to the wand. In a study involving 210 patients, RF detection systems had 100% sensitivity and specificity for detecting retained sponges, even in morbidly obese patients. To minimize human error during the scanning process, a new system has been developed where the patient lies on a mat that emits the RF signal and serves as the antenna. In a separate study involving 203 patients, the system using the RF mat had a sensitivity and specificity of 98.5% and 100% respectively. To carry out these studies, sponges were placed underneath patients in a blinded manner. Although it would seem that RF detection systems are superior to manual counting, it should be emphasized that an RF detection system is not a substitute for manual counting, but rather an adjunct to manual counting. The RF Assure ® Detection System manufactured by RF Surgical System Inc. is now routinely used to detect retained sponges in the operating rooms at the University of California, San Diego (UCSD) and at ~200 other hospitals.

There has been some concern that the radio frequency signals emitted by the RF Assure® device might reprogram the settings on CIED such as pacemakers and internal cardiac defibrillators (ICDs), since CIEDs are commonly programmed remotely using RF signals. For this reason, the technology is not commonly used in surgeries with patients with CIEDs. In vitro testing was done in a controlled setting to assess the compatibility of the RF Surgical Detection Technology with permanent pacemakers and ICDs as well as temporary pacemakers, specifically the Boston Scientific Cognis, St. Jude Medical Promote, and Medtronic Virtuoso devices, finding no interference between the RF Assure® system and these devices. Although anecdotal experience and in vitro testing has indicated that the RF Assure® system does not affect CIEDs, it has not been established whether the RF Assure® system is suitable to use in patients with CIEDs in a clinical setting. The purpose of this study is to determine whether the RF Assure® device actually can reprogram CIEDs or cause them to malfunction when used in routine clinical practice. This question is of particular importance because of the rising number of surgical patients, especially cardiac patients, that have an implanted CIED.

Being tested
RF Assure Scanning
Keywords
CIED, pacemaker, implantable cardioverter defibrillator, ICD, radiofrequency, cardiac device programming, RF sponge detection, RF assure, cardiac internal electronic devices, temporary pacing, pulmonary thromboendarterectomy, pacing interference, Arrhythmia
Medical subject headings
Cardiomyopathies, Arrhythmias, Cardiac, Heart Failure

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • RF Assure ScanningDevice

    CIED and temporary pacemaker patients will be scanned with RF surgical sponge detection wand and mat as well as with and without the RF sponge. The scanning will only be conducted on the device scheduled for removal. Patients will have device interrogated before and after scanning (for CIEDs) to determine if there have been any clinically significant changes in programming or settings due to scanning with RF technology. Vitals will be monitored and recorded before, during, and after scanning.

Study groups in NCT02111980
Study groupRoleWhat they receive
RF Assure ScanningExperimentalThe patient's CIED will be interrogated prior to the study to obtain a baseline reading. The patient will be asked to lie down on the RF Assure® Detection Mat with a sponge placed underneath his or her shoulder. The RF Assure® mat and wand will be activated to detect the sponge. The sponge will be removed from underneath the patient's shoulder, and the RF system will be re-activated to obtain a clear reading. The patient's CIED will be re-interrogated to determine if the RF Assure system caused any changes to the CIED parameters or function.
Receives: Device: RF Assure Scanning

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The Pacing Mode on the CIED was measured prior to RF scanning (with sponge) and after the scanning (after sponge removal) in order to evaluate if any significant changes in the pacing mode setting resulted. The following CIED modes were evaluated: DDD (dual chamber pacing, sensing, triggered and inhibited mode), VVI (ventricular pacing, sensing, and inhibited mode), DDI (dual pacing, sensing, and inhibited mode), AAI (atrial pacing, sensing, and inhibited mode). The number of patients' device mode switched between these settings was tabulated and is shown in the below table. Multiple post-scan assessments were not made.

    Measured: Baseline and 15 minutes

  2. Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The patient's pacing polarity was measured prior to scanning with sponge and after scanning with sponge via device interrogation. The following pacing polarity measurements were evaluated: right atrium/right ventricle (RA/RV) bipolar polarity, left ventricle (LV) bipolar polarity, and left ventricle (LV) unipolar polarity. Note that not all study patients had LV leads implanted. Multiple post-scan assessments were not made.

    Measured: Baseline and 15 minutes

  3. Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure

    The base rate on patients' devices was measured before scanning with sponge and after RF scanning with sponge. The median was determined and is presented below with standard deviation for both times. Multiple post-scan assessments were not made.

    Measured: Baseline and 15 minutes

  4. Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The max tracking rate on the CIEDs was measured prior to scanning with the sponge and post scanning with sponge. The max tracking rate is the maximum atrial rate at which a pacemaker will deliver a ventricular pacing stimulus following each sensed atrial beat. Below, the median and standard deviation are presented. Please note that multiple post-scan assessments were not made for any patient.

    Measured: Baseline and 15 minutes

  5. Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The atrio-ventricular delay was measured prior to scanning with sponge and post RF scanning with sponge. Please note that multiple post-scan assessments were not made for any of the patients presented here.

    Measured: Baseline and 15 minutes

  6. Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    Of the patients enrolled, battery capacity changes were measured prior to and post RF scanning with sponge. Multiple post-scan assessments were not made for any of the patients presented here.

    Measured: Baseline and 15 minutes

  7. Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The impedance values of right ventricle (RV), right atrium (RA), and left ventricle (LV) leads was measured prior to RF scanning with sponge and post RF scanning with sponge. Multiple post-scan assessments were not made for any patient represented here.

    Measured: Baseline and 15 minutes

  8. Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The shock impedance changes were measured (right ventricle (RV) coil and superior vena cava (SVC) coil) prior to and post RF scanning with sponge. This was measured by performing a device interrogation in the electrophysiology (EP) lab. Multiple post-scan measurements were not taken for any of the patients presented here.

    Measured: Baseline and 15 minutes

  9. Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The P and R waves were measured via device interrogation prior to and post RF scanning with sponge. Multiple post-scan measurements were not made for any of the participants represented here.

    Measured: Baseline and 15 minutes

  10. Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    Via device interrogation, capture threshold changes (the minimum amount of electricity that the box has to emit to pace the heart) were measured prior to and post RF Scanning with sponge. Multiple post-scan measurements were not taken for any participants presented here.

    Measured: Baseline and 15 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 Apr 2014 (estimated)
  2. Study startMar 2014
  3. Primary completionMar 2015
  4. Study completionMay 2015

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
50 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02111980
Sponsor's protocol ID
131198
Acronym
RF
Status last verified
Dec 2019
First submitted
9 Apr 2014
Last updated
24 Dec 2019
Data monitoring committee
No

Publications

  • Cima RR, Kollengode A, Garnatz J, Storsveen A, Weisbrod C, Deschamps C. Incidence and characteristics of potential and actual retained foreign object events in surgical patients. J Am Coll Surg. 2008 Jul;207(1):80-7. doi: 10.1016/j.jamcollsurg.2007.12.047. Epub 2008 May 23.

    PubMed 18589366
  • Egorova NN, Moskowitz A, Gelijns A, Weinberg A, Curty J, Rabin-Fastman B, Kaplan H, Cooper M, Fowler D, Emond JC, Greco G. Managing the prevention of retained surgical instruments: what is the value of counting? Ann Surg. 2008 Jan;247(1):13-8. doi: 10.1097/SLA.0b013e3180f633be.

    PubMed 18156916
  • Gawande AA, Studdert DM, Orav EJ, Brennan TA, Zinner MJ. Risk factors for retained instruments and sponges after surgery. N Engl J Med. 2003 Jan 16;348(3):229-35. doi: 10.1056/NEJMsa021721.

    PubMed 12529464
  • Steelman VM, Alasagheirin MH. Assessment of radiofrequency device sensitivity for the detection of retained surgical sponges in patients with morbid obesity. Arch Surg. 2012 Oct;147(10):955-60. doi: 10.1001/archsurg.2012.1556.

    PubMed 23070411
  • Steelman VM. Sensitivity of detection of radiofrequency surgical sponges: a prospective, cross-over study. Am J Surg. 2011 Feb;201(2):233-7. doi: 10.1016/j.amjsurg.2010.05.001.

    PubMed 21266216

Questions about this trial

What is this clinical trial studying?

A study testing RF Assure Scanning in people with Cardiomyopathy and Cardiac Arrhythmia, run by University of California, San Diego at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 24 Dec 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 24 Dec 2019). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure (time frame: Baseline and 15 minutes). The study lists 10 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT02111980, which also lists 5 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Dec 2019.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in