Cardiac Resynchronization TherapyNAVIGATE X4NCT02071173
Evaluation of ACUITY™ X4 Quadripolar Coronary Venous Leads and RELIANCE™ 4-FRONT Defibrillation Leads
A study testing ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation lead in people with Cardiac Resynchronization Therapy, run by Boston Scientific Corporation at 96 sites in United States.
Trial summary
The objective of the NAVIGATE X4 Clinical Study is to gather data to establish the safety, performance and effectiveness of the ACUITY™ X4 quadripolar coronary venous leads and the RELIANCE 4-FRONT™ ventricular defibrillation leads to satisfy FDA requirements for pre-market submission. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Cardiac Resynchronization Therapy
- What is being tested?
- ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation lead
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 2,244 participants
- Where
- 96 sitesUnited States
- Study dates
- 14 Apr 2014to 6 Oct 2020
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Subjects indicated for a CRT-D that fulfill one of the following 5 criteria[1]:
- Subject with left ventricular ejection fraction (LVEF) less than or equal to 35%, sinus rhythm, left bundle branch block (LBBB) with a QRS duration greater than or equal to 150 ms, and a NYHA class II, III or ambulatory IV symptoms on Guideline-Directed Medical Therapy (GDMT)*
- Subject with LVEF less than or equal to 35%, sinus rhythm, LBBB with a QRS duration 120 to 149 ms, and NYHA class II, III, or ambulatory IV symptoms on GDMT*
- Subject with LVEF less than or equal to 35%, sinus rhythm, a non-LBBB pattern with a QRS duration greater than or equal to 150 ms, and NYHA class III/ ambulatory class IV symptoms on GDMT*
- Subject with atrial fibrillation and LVEF less than or equal to 35% on GDMT* if a) the subject requires ventricular pacing or otherwise meets CRT criteria [listed here] and b) AV nodal ablation or pharmacologic rate control will allow near 100% ventricular pacing with CRT
- Subject on GDMT* who have LVEF less than or equal to 35% and are undergoing new device placement with anticipated requirement for significant (>40%) ventricular pacing *GDMT = Guideline-directed medical therapy (formerly known as optimal pharmaceutical therapy (OPT)), represents optimal medical therapy as defined by ACCF/ AHA guideline-recommended therapies (primarily Class I)
Show all 9Show fewer
- Subject is intended to receive the ACUITY X4 LV lead and RELIANCE 4-FRONT RV lead (optional in Study Phase 1) and BSC CRT-D with quadripolar header as their initial (de novo) cardiac implants
- Subject is willing and capable of providing informed consent (which can include the use of a legally authorized representative [LAR] for documentation of informed consent) and participating in all testing associated with this investigation at an approved center and at the intervals defined by this protocol
- Subject is age 18 or above, or of legal age to give informed consent specific to state and national law
Who cannot join
- Subjects who meet any one of the following criteria will be excluded from this clinical study.
- Subject has a known or suspected sensitivity to dexamethasone acetate (DXA)
- Subject has a mechanical tricuspid heart valve
- Subject is enrolled in any other concurrent study, with the exception of local mandatory governmental registries and observational studies/registries* that are not in conflict and do not affect the following:
- Schedule of procedures for the Study (i.e. should not cause additional or missed visits);
- Study outcome (i.e. involve medications that could affect the heart rate of the subject);
Show all 11Show fewer
- Conduct of the Study per good clinical practice (GCP) / International Organization for Standardization (ISO) 14155:2011/ 21 Code of Federal Regulations (CFR) 812, local regulations
- Subject is currently on the active heart transplant list
- Subject has a documented life expectancy of less than twelve months
- Women of childbearing potential who are or might be pregnant at the time of study enrollment or CRT-D System implant (method of assessment upon physician's discretion)
- Subjects currently requiring chronic dialysis *Sponsors of such studies/registries should be informed and Boston Scientific must be informed by the investigator about the parallel conduct of these projects in the subject and of the project's basic nature. The decision if a desired mandatory governmental registry or observational study/ registry is in conflict with this exclusion criterion is up to the enrolling investigator.
Trial locations and contacts
96 sites in 1 country.
96 sites in 1 country
United States96 sites
Abington Memorial Hospital
Abington, Pennsylvania, 19001
Lehigh Valley Hospital
Allentown, Pennsylvania, 18103
Providence Alaska Medical Center
Anchorage, Alaska, 99508
Emory University Hospital
Atlanta, Georgia, 30308
Texas Cardiac Arrhythmia Research
Austin, Texas, 78705
Heart Hospital of Austin
Austin, Texas, 78756
Union Memorial Hospital
Baltimore, Maryland, 21237
Eastern Maine Medical Center
Bangor, Maine, 04401
SouthEast Texas Clinical Research Center
Beaumont, Texas, 77702
PeaceHealth St. Joseph Medical Center
Bellingham, Washington, 98225
University of Alabama at Birmingham
Birmingham, Alabama, 35294
Beth Israel Deaconess Medical Center
Boston, Massachusetts, 02215
Maimonides Medical Center
Brooklyn, New York, 11219
Deborah Heart and Lung Center
Browns Mills, New Jersey, 08015
Aultman Hospital
Canton, Ohio, 44710
University of North Carolina Hospital
Chapel Hill, North Carolina, 27599
Medical University of South Carolina
Charleston, South Carolina, 29425
Charleston Area Medical Center
Charleston, West Virginia, 25304
University of Chicago
Chicago, Illinois, 60637
Good Samaritan Hospital
Cincinnati, Ohio, 45220
Cleveland Clinic Foundation
Cleveland, Ohio, 44195
Boone Hospital Center
Columbia, Missouri, 65201
Ohio State University Medical Center
Columbus, Ohio, 43210
OhioHealth Research and Innovation Institute - Riverside Methodist Hospital
Columbus, Ohio, 43214
Medical City Dallas Hospital
Dallas, Texas, 75230
Walnut Hill Medical Center
Dallas, Texas, 75231
St. Mary's Duluth Clinic Regional Heart Center
Duluth, Minnesota, 55805
Duke University Medical Center
Durham, North Carolina, 27710
Porter Memorial Hospital
Dyer, Indiana, 46311
Alexian Brothers Medical Center
Elk Grove Village, Illinois, 60007
Providence Regional Medical Center Everett
Everett, Washington, 98201
PeeDee Cardiology Associates PA
Florence, South Carolina, 29506
Delray Medical Center
Fort Lauderdale, Florida, 33334
Lutheran Medical Group
Fort Wayne, Indiana, 48804
Plaza Medical Center of Fort Worth
Fort Worth, Texas, 76104
Stern Cardiovascular Foundation, Inc.
Germantown, Tennessee, 38138
Spectrum Health Hospitals
Grand Rapids, Michigan, 49503
Cooper Hospital - University Medical Center
Haddon Heights, New Jersey, 08035
Hartford Hospital
Hartford, Connecticut, 06102
Huntsville Hospital
Huntsville, Alabama, 35801
St. Vincent's Hospital
Indianapolis, Indiana, 46260
University of Iowa Hospitals and Clinics
Iowa City, Iowa, 52242
Cardiology Associates of Northeast Arkansas, P.A.
Jonesboro, Arkansas, 72401
University of Kansas Hospital
Kansas City, Kansas, 66160
St. Luke's Hospital of Kansas City
Kansas City, Missouri, 64111
Lakeland Regional Medical Center
Lakeland, Florida, 38805
West Florida Cardiology Network, LLC
Largo, Florida, 33770
Gwinnett Hospital System Inc.
Lawrenceville, Georgia, 30046
Baptist Health Lexington
Lexington, Kentucky, 40503
Nebraska Heart Institute
Lincoln, Nebraska, 68526
Kaiser Permanente
Los Angeles, California, 90027
USC Medical Center
Los Angeles, California, 90033
University of Louisville Hospital
Louisville, Kentucky, 40202
Bon Secours Heart & Vascular Institute
Mechanicsville, Virginia, 23116
Asante Rogue Regional Medical Center
Medford, Oregon, 97504
Banner Baywood Heart Hospital
Mesa, Arizona, 85206
Monongalia General Hospital
Morgantown, West Virginia, 26505
Ball Memorial Hospital
Muncie, Indiana, 47303
Intermountain Medical Center
Murray, Utah, 84107
St. Thomas Research Institute, LLC
Nashville, Tennessee, 37205
Vanderbilt University Medical Center
Nashville, Tennessee, 37215
Jersey Shore University Medical Center
Neptune City, New Jersey, 07753
Yale University School of Medicine
New Haven, Connecticut, 06510
Long Island Jewish Medical Center
New Hyde Park, New York, 11040
Columbia University Medical Center
New York, New York, 10032
Sentara Norfolk General Hospital
Norfolk, Virginia, 23507
Alta Bates Medical Center
Oakland, California, 94609
Providence St. Vincent Medical Center
Portland, Oregon, 97225
Rapid City Regional Hospital
Rapid City, South Dakota, 57701
Virginia Cardiovascular Specialist
Richmond, Virginia, 23225
Valley Hospital
Ridgewood, New Jersey, 07450
Mercy General Hospital
Sacramento, California, 95819
St. Cloud Hospital
Saint Cloud, Minnesota, 56303
HealthEast St. Joseph's Hospital
Saint Paul, Minnesota, 55102
United Heart and Vascular Clinic
Saint Paul, Minnesota, 55102
Sharp Memorial Hospital
San Diego, California, 92123
Scottsdale Healthcare - Osborn
Scottsdale, Arizona, 85258
Avera Heart Hospital of South Dakota
Sioux Falls, South Dakota, 57108
Spartanburg Regional Medical Center
Spartanburg, South Carolina, 29303
Kootenai Heart Clinics
Spokane, Washington, 99204
Providence Health & Services - Washington
Spokane, Washington, 99204
St. John's Hospital
Springfield, Illinois, 62769
Cox Health
Springfield, Missouri, 65807
Barnes Jewish Hospital
St Louis, Missouri, 63110
Wheeling Hospital Inc.
Steubenville, Ohio, 43952
Stony Brook University Hospital
Stony Brook, New York, 11794
Tallahassee Memorial Hospital
Tallahassee, Florida, 32308
Tampa General Hospital
Tampa, Florida, 33606
Montefiore Medical Center
The Bronx, New York, 10467
University of Texas Houston Health Science Center
The Woodlands, Texas, 77384
The Toledo Hospital
Toledo, Ohio, 43615
Trinity Mother Frances Health System
Tyler, Texas, 75701
Buffalo General Hospital
Williamsville, New York, 14221
Mercy Hospital of Buffalo, Catholic Health System
Williamsville, New York, 14221
York Hospital
York, Pennsylvania, 17403
St. Joseph Mercy Hospital
Ypsilanti, Michigan, 48197
About this study
Official title: Evaluation of ACUITY™ X4 Quadripolar Coronary Venous Leads and RELIANCE™ 4-FRONT Defibrillation Leads: NAVIGATE X4
The objective of the NAVIGATE X4 Clinical Study is to gather data to establish the safety, performance and effectiveness of the ACUITY™ X4 quadripolar coronary venous leads and the RELIANCE 4-FRONT™ ventricular defibrillation leads to satisfy FDA requirements for pre-market submission. Additionally, data from this study will be used to support post-market approval requirements for the ACUITY X4 and RELIANCE 4-FRONT leads.
- Conditions
- Cardiac Resynchronization Therapy
- Being tested
- ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation lead
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation leadDevice
Also called LV Quadripolar coronary venous lead, RV defibrillation Lead
The ACUITY X4 leads are intended for chronic left ventricular pacing and sensing. The leads have 3 tip configuration designs (straight tip, short tip spiral, and long tip spiral), which is intended to provide choices for patients with different anatomies.
The RELIANCE 4-FRONT leads are designed for right-side heart applications within the ventricle, atrium, and superior vena cava. It is intended for permanent sensing, pacing, and defibrillation when used with a compatible implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy defibrillator (CRT-D).
Subjects were allowed to receive both the ACUITY X4 and RELIANCE 4-FRONT lead.
| Study group | Role | What they receive |
|---|---|---|
| Enrolled Patients | Experimental | Subjects undergo an implant procedure to receive at least one investigational lead -- ACUITY X4 left ventricular (LV) CRT lead, RELIANCE 4-FRONT right ventricular (RV) ICD lead Receives: Device: ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation lead |
What is being measured
The primary outcome is the main question the study is designed to answer.
Percentage of ACUITY X4 Spiral Leads Free From Complication Through 6 Months
Lead-related complication-free rate of ACUITY X4 Spiral Leads from implant through 6 months post implant.
Measured: Implant through 6 months
Percentage of ACUITY X4 Straight Leads Free From Complication Through 6 Months
Lead-related complication-free rate of ACUITY X4 Straight Leads from implant through 6 months post implant.
Measured: Implant through 6 months
Percentage of ACUITY X4 Leads With Acceptable Pacing Capture Thresholds (PCT)
PCT measurements in the programmed configuration were collected at 3 months post-implant. Programmed configuration refers to the pacing configuration that was selected by the physician to provide LV pacing therapy. The percentage of measurements with PCT ≤ 2.5 V were calculated.
Measured: Implant through 3 months
Percentage of ACUITY X4 Spiral Leads With Acceptable Pacing Capture Thresholds (PCT) Using the Best Proximal Electrode
PCT measurements in the in the proximal zone (electrodes 2, 3, 4) were collected at 3 months post-implant. Physician investigators were instructed to use the best proximal electrode (E2, E3 or E4) as the cathode and the RV lead coil or pulse generator as the anode; best was defined as the electrode with the lowest PCT without phrenic nerve stimulation. The percentage of measurements with PCT ≤ 2.5 V were calculated.
Measured: 3 months
Percentage of RELIANCE 4-FRONT Leads Free From Complication Through 3 Months
Lead-related complication-free rate of RELIANCE 4-FRONT Leads from implant through 3 months post implant.
Measured: Implant through 3 months
Percentage of RELIANCE 4-FRONT Leads Free From Complication From 3 Through 24 Months
Lead-related complication-free rate of RELIANCE 4-FRONT Leads from 3 through 24 months post implant.
Measured: 3 months through 24 months
Mean Pacing Capture Threshold (PCT) of RELIANCE 4-FRONT Leads at 3 Months
PCT measurements from RELIANCE 4-FRONT leads were collected at 3 months post-implant. Measurements were performed using 0.5 millisecond pulse width.
Measured: Implant through 3 months
Secondary outcomes (9)
Mean Sensed Amplitude of ACUITY X4 Spiral Leads at 3 Months
Sensed amplitude measurements from ACUITY X4 Spiral leads were collected at 3 months post-implant.
Measured: 3 months
Mean Sensed Amplitude of ACUITY X4 Straight Leads at 3 Months
Sensed amplitude measurements from ACUITY X4 Straight leads were collected at 3 months post-implant.
Measured: 3 months
Mean Pacing Impedance of ACUITY X4 Spiral Leads at 3 Months
Pacing impedance measurements from ACUITY X4 Spiral leads were collected at 3 months post-implant.
Measured: 3 months
Mean Pacing Impedance ACUITY X4 Straight Leads at 3 Months
Pacing impedance measurements from ACUITY X4 Straight leads were collected at 3 months post-implant.
Measured: 3 months
Mean Detection Time of Induced Ventricular Tachycardia / Ventricular Fibrillation (VT/VF) Episodes Using RELIANCE 4-FRONT Leads
The mean detection time, in seconds, of induced VT/VF episodes occurring within 30 days of implant were evaluated. Subjects were required to implanted with RELIANCE 4-FRONT leads to be evaluated.
Measured: Within 30 days of implant
Mean Sensed Amplitude of RELIANCE 4-FRONT Leads at 3 Months
Sensed amplitude measurements from RELIANCE 4-FRONT leads were collected at 3 months post-implant.
Measured: 3 months
Mean Pacing Impedance of RELIANCE 4-FRONT Active Fixation Leads at 3 Months
Pacing impedance measurements from RELIANCE 4-FRONT Active Fixation leads were collected at 3 months post-implant.
Measured: 3 months
Mean Pacing Impedance of RELIANCE 4-FRONT Passive Fixation Leads at 3 Months
Pacing impedance measurements from RELIANCE 4-FRONT Passive Fixation leads were collected at 3 months post-implant.
Measured: 3 months
Percentage of Successfully Converted Induced Ventricular Tachycardia / Ventricular Fibrillation (VT/VF) Episodes Using RELIANCE 4-FRONT Leads
Ventricular Tachyarrhythmia (VT/VF) Shock Conversion Efficacy, analyzed within 30 days of implant
Measured: Within 30 days of implant
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted25 Feb 2014 (estimated)
- Study start14 Apr 2014
- Primary completion19 Apr 2017
- Study completion6 Oct 2020
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Cardiac Resynchronization TherapyStudies of the same condition, open and under way.
- More trials run by Boston Scientific CorporationThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Model description
- Every subject had access to two study devices and were followed as a single group.
- Masking
- None (open label)
- Enrolment
- 2,244 participants (actual)
Sponsor and investigators
- Lead sponsor
- Boston Scientific Corporation
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Suneet Mittal, MD, Principal investigator, The Valley HospitalMartin Burke, DO, Principal investigator, University of Chicago
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02071173
- Sponsor's protocol ID
- C1481
- Acronym
- NAVIGATE X4
- Status last verified
- Dec 2020
- First submitted
- 21 Feb 2014
- Last updated
- 7 Jan 2021
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Publications
Mittal S, Nair D, Padanilam BJ, Ciuffo A, Gupta N, Gallagher P, Goldner B, Hammill EF, Wold N, Stein K, Burke M. Performance of Anatomically Designed Quadripolar Left Ventricular Leads: Results from the NAVIGATE X4 Clinical Trial. J Cardiovasc Electrophysiol. 2016 Oct;27(10):1199-1205. doi: 10.1111/jce.13044. Epub 2016 Aug 19.
PubMed 27434039
Study at a glance
From the registry record, last updated by the sponsor on 7 Jan 2021. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Evaluation of ACUITY™ X4 Quadripolar Coronary Venous Leads and RELIANCE™ 4-FRONT Defibrillation Leads: NAVIGATE X4
- ClinicalTrials.gov ID
- NCT02071173
- Acronym / sponsor ID
- NAVIGATE X4 / C1481
- Conditions
- Cardiac Resynchronization Therapy
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 2,244 participants (actual)
- Lead sponsor
- Boston Scientific CorporationIndustry (pharma, biotech, devices)
- Study start
- 14 Apr 2014 (actual)
- Primary completion
- 19 Apr 2017 (actual)
- Study completion
- 6 Oct 2020 (actual)
- First posted
- 25 Feb 2014 (estimated)
- Results
- Posted 30 Oct 2020
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The objective of the NAVIGATE X4 Clinical Study is to gather data to establish the safety, performance and effectiveness of the ACUITY™ X4 quadripolar coronary venous leads and the RELIANCE 4-FRONT™ ventricular defibrillation leads to satisfy FDA requirements for pre-market submission. Additionally, data from this study will be used to support post-market approval requirements for the ACUITY X4 and RELIANCE 4-FRONT leads.
Study design
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
- Model description
- Every subject had access to two study devices and were followed as a single group.
| Arm | Type | Receives |
|---|---|---|
| Enrolled Patients | Experimental | Device: ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation leadSubjects undergo an implant procedure to receive at least one investigational lead -- ACUITY X4 left ventricular (LV) CRT lead, RELIANCE 4-FRONT right ventricular (RV) ICD lead |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (9 listed)
- Subject with left ventricular ejection fraction (LVEF) less than or equal to 35%, sinus rhythm, left bundle branch block (LBBB) with a QRS duration greater than or equal to 150 ms, and a NYHA class II, III or ambulatory IV symptoms on Guideline-Directed Medical Therapy (GDMT)*
- Subject with LVEF less than or equal to 35%, sinus rhythm, LBBB with a QRS duration 120 to 149 ms, and NYHA class II, III, or ambulatory IV symptoms on GDMT*
- Subject with LVEF less than or equal to 35%, sinus rhythm, a non-LBBB pattern with a QRS duration greater than or equal to 150 ms, and NYHA class III/ ambulatory class IV symptoms on GDMT*
- Subject with atrial fibrillation and LVEF less than or equal to 35% on GDMT* if a) the subject requires ventricular pacing or otherwise meets CRT criteria [listed here] and b) AV nodal ablation or pharmacologic rate control will allow near 100% ventricular pacing with CRT
- Subject on GDMT* who have LVEF less than or equal to 35% and are undergoing new device placement with anticipated requirement for significant (>40%) ventricular pacing *GDMT = Guideline-directed medical therapy (formerly known as optimal pharmaceutical therapy (OPT)), represents optimal medical therapy as defined by ACCF/ AHA guideline-recommended therapies (primarily Class I)
Key exclusion criteria (11 listed)
- Subjects who meet any one of the following criteria will be excluded from this clinical study.
- Subject has a known or suspected sensitivity to dexamethasone acetate (DXA)
- Subject has a mechanical tricuspid heart valve
- Subject is enrolled in any other concurrent study, with the exception of local mandatory governmental registries and observational studies/registries* that are not in conflict and do not affect the following:
- Schedule of procedures for the Study (i.e. should not cause additional or missed visits);
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation leadAlso: LV Quadripolar coronary venous lead, RV defibrillation LeadThe ACUITY X4 leads are intended for chronic left ventricular pacing and sensing. The leads have 3 tip configuration designs (straight tip, short tip spiral, and long tip spiral), which is intended to provide choices for patients with different anatomies. The RELIANCE 4-FRONT leads are designed for right-side heart applications within the ventricle, atrium, and superior vena cava. It is intended for permanent sensing, pacing, and defibrillation when used with a compatible implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy defibrillator (CRT-D). Subjects were allowed to receive both the ACUITY X4 and RELIANCE 4-FRONT lead. | Device | Enrolled Patients |
Primary outcome
Percentage of ACUITY X4 Spiral Leads Free From Complication Through 6 Months
Time frame: Implant through 6 months
Lead-related complication-free rate of ACUITY X4 Spiral Leads from implant through 6 months post implant.
Percentage of ACUITY X4 Straight Leads Free From Complication Through 6 Months
Time frame: Implant through 6 months
Lead-related complication-free rate of ACUITY X4 Straight Leads from implant through 6 months post implant.
Percentage of ACUITY X4 Leads With Acceptable Pacing Capture Thresholds (PCT)
Time frame: Implant through 3 months
PCT measurements in the programmed configuration were collected at 3 months post-implant. Programmed configuration refers to the pacing configuration that was selected by the physician to provide LV pacing therapy. The percentage of measurements with PCT ≤ 2.5 V were calculated.
Percentage of ACUITY X4 Spiral Leads With Acceptable Pacing Capture Thresholds (PCT) Using the Best Proximal Electrode
Time frame: 3 months
PCT measurements in the in the proximal zone (electrodes 2, 3, 4) were collected at 3 months post-implant. Physician investigators were instructed to use the best proximal electrode (E2, E3 or E4) as the cathode and the RV lead coil or pulse generator as the anode; best was defined as the electrode with the lowest PCT without phrenic nerve stimulation. The percentage of measurements with PCT ≤ 2.5 V were calculated.
Percentage of RELIANCE 4-FRONT Leads Free From Complication Through 3 Months
Time frame: Implant through 3 months
Lead-related complication-free rate of RELIANCE 4-FRONT Leads from implant through 3 months post implant.
Percentage of RELIANCE 4-FRONT Leads Free From Complication From 3 Through 24 Months
Time frame: 3 months through 24 months
Lead-related complication-free rate of RELIANCE 4-FRONT Leads from 3 through 24 months post implant.
Mean Pacing Capture Threshold (PCT) of RELIANCE 4-FRONT Leads at 3 Months
Time frame: Implant through 3 months
PCT measurements from RELIANCE 4-FRONT leads were collected at 3 months post-implant. Measurements were performed using 0.5 millisecond pulse width.
9 secondary outcomes
Mean Sensed Amplitude of ACUITY X4 Spiral Leads at 3 Months
Time frame: 3 months
Mean Sensed Amplitude of ACUITY X4 Straight Leads at 3 Months
Time frame: 3 months
Mean Pacing Impedance of ACUITY X4 Spiral Leads at 3 Months
Time frame: 3 months
Mean Pacing Impedance ACUITY X4 Straight Leads at 3 Months
Time frame: 3 months
Mean Detection Time of Induced Ventricular Tachycardia / Ventricular Fibrillation (VT/VF) Episodes Using RELIANCE 4-FRONT Leads
Time frame: Within 30 days of implant
Mean Sensed Amplitude of RELIANCE 4-FRONT Leads at 3 Months
Time frame: 3 months
Mean Pacing Impedance of RELIANCE 4-FRONT Active Fixation Leads at 3 Months
Time frame: 3 months
Mean Pacing Impedance of RELIANCE 4-FRONT Passive Fixation Leads at 3 Months
Time frame: 3 months
Percentage of Successfully Converted Induced Ventricular Tachycardia / Ventricular Fibrillation (VT/VF) Episodes Using RELIANCE 4-FRONT Leads
Time frame: Within 30 days of implant
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 7 Jan 2021.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Derived from this study: PubMed 27434039Mittal S, Nair D, Padanilam BJ, Ciuffo A, Gupta N, Gallagher P, Goldner B, Hammill EF, Wold N, Stein K, Burke M. Performance of Anatomically Designed Quadripolar Left Ventricular Leads: Results from the NAVIGATE X4 Clinical Trial. J Cardiovasc Electrophysiol. 2016 Oct;27(10):1199-1205. doi: 10.1111/jce.13044. Epub 2016 Aug 19.
Related trials for Cardiac Resynchronization Therapy
Recruiting
Safety and Effectiveness of Left Bundle Branch Area Pacing Versus Conventional Cardiac Resynchronization Therapy in Heart Failure
Heart Failure - NYHA II - IV
RecruitingObservational study
BIO|STREAM.CSP Study
Cardiac Pacing, Cardiac Pacemaker
Active, not recruiting
BIO|MASTER.CSP Study
Cardiac Pacing, Artificial, Cardiac Pacemaker, Artificial
Active, not recruiting
Cardiac Resynchronization Therapy in Previously Untreatable and High Risk Upgrade Patients
Heart Failure
Recruiting
Bachmann's Bundle Versus Right Atrial Appendage Pacing
Atrial Fibrillation (AF)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Suneet Mittal, MD principal investigator
The Valley Hospital
Martin Burke, DO principal investigator
University of Chicago
96 sites in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing ACUITY X4 quadripolar coronary venous lead and RELIANCE 4-FRONT defibrillation lead in people with Cardiac Resynchronization Therapy, run by Boston Scientific Corporation at 96 sites in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 9 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 7 Jan 2021. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 96 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 30 Oct 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Percentage of ACUITY X4 Spiral Leads Free From Complication Through 6 Months (time frame: Implant through 6 months). The study lists 7 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT02071173, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 7 Jan 2021.