A Clinical Assessment of the Hemacopolymer Daily Disposable Limbal Ring Soft Contact Lens

A study testing Hema-copolymer Lens and etafilcon A Lens in people with Myopia, run by CooperVision, Inc. at 1 site in United Kingdom.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This purpose of this study is to compare the clinical performance and subjective acceptance of two different daily disposable limbal ring lenses. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Myopia
What is being tested?
Hema-copolymer Lens and etafilcon A Lens
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
30 participants
Where
1 siteUnited Kingdom
Study dates
Apr 2014to Oct 2014
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subjects will only be eligible for the study if:
  • They are 18 years of age and above.
  • They understand their rights as a research subject and are willing and able to sign a Statement of Informed Consent.
  • They are willing and able to follow the protocol.
  • They agree not to participate in other clinical research for the duration of this study.
  • They can attain at least 6/9 in each eye with the study lenses within the available power range.
  • They can be fitted with study lenses within the available power range.
  • The currently use soft contact lenses, or have done so within the last six months.

Who cannot join

  • Subjects will not be able to take part in the study if:
  • They have an ocular disorder which would normally contraindicate contact lens wear.
  • They have a systemic disorder which would normally contraindicate contact lens wear.
  • They are using any topical medication such as eye drops or ointment.
  • They have had cataract surgery.
  • They have had corneal refractive surgery.
Show all 11Show fewer
  • They have any corneal distortion resulting from previous hard or rigid lens wear or have keratoconus.
  • They are pregnant or lactating.
  • They have any ocular abnormality which would, in the opinion of the investigator, normally contraindicate contact lens wear.
  • They have any infectious disease which would normally contraindicate contact lens wear, or may, in the opinion of the investigator, pose a risk to study personnel; or any immunosuppressive disease (e.g. HIV), or a history of anaphylaxis or severe allergic reactions.
  • They have taken part in any other clinical trial or research, within two weeks prior to starting this study.

Trial locations and contacts

1 site in 1 country.

United Kingdom1 site
  • Eurolens Research - The University of Manchester

    Manchester, M13 9PL

Other trials in United Kingdom

About this study

This purpose of this study is to compare the clinical performance and subjective acceptance of two different daily disposable limbal ring lenses.

Read the detailed description

This is a subject-masked, dispensing study designed to evaluate the clinical performance and subjective acceptance when comparing two different daily disposable limbal ring lenses. Limbal ring lenses, which as well as correcting vision, have a colored ring which enhance the wearer's iris and are particularly popular in East Asian countries.

Conditions
Myopia
Being tested
Hema-copolymer Lens, etafilcon A Lens
Medical subject headings
Myopia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Hema-copolymer LensDevice

    contact lens

  • etafilcon A LensDevice

    contact lens

Study groups in NCT01965262
Study groupRoleWhat they receive
Hema-copolymer LensActive comparatorParticipants were randomized to wear the Hema-copolymer lens pair for one week during the cross over study.
Receives: Device: Hema-copolymer Lens; Device: etafilcon A Lens
etafilcon A LensActive comparatorParticipants were randomized to wear the etafilcon A lens pair for one week during the cross over study.
Receives: Device: Hema-copolymer Lens; Device: etafilcon A Lens

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Visual Acuity - Hema-copolymer and Etafilcon A

    Visual acuity measured by logMAR of hema-copolymer and etafilcon A lenses assessed at baseline.

    Measured: Baseline

  2. Visual Acuity - Hema-copolymer and Etafilcon A

    Visual acuity measured by logMAR for hema-copolymer and etafilcon A lenses assessed at 1 week.

    Measured: 1 week

  3. Biomicroscopy - Hema-copolymer and Etafilcon A

    Biomicroscopy is analyzed for hema-copolymer and etafilcon A at 1 week. (Scale 0-4, 0=normal, 4=severe).

    Measured: 1 week

  4. Lens Surface - Deposition - Hema-copolymer and Etafilcon A

    Lens surface of deposition for hema-copolymer and etafilcon A pair of lenses assessed at baseline. (Each pair of lenses worn by the participant was assigned a single grade). Overall score measured by Grade 0-4; 0=absent, clean surface, 4= multiple deposits.

    Measured: Baseline

  5. Lens Surface - Deposition - Hema-copolymer and Etafilcon A

    Lens surface of deposition for hema-copolymer and etafilcon A pair of lenses assessed at 1 week. (Each pair of lenses worn by the participant was assigned a single grade). Overall score measured by Grade 0-4; 0=absent, clean surface, 4= multiple deposits.

    Measured: 1 week

  6. Lens Surface - Debris - Hema-copolymer and Etafilcon A

    Lens surface of debris for hema-copolymer and etafilcon A pair of lenses assessed at baseline. (Each pair of lenses worn by the participant was assigned a single grade). Overall score measured by Grade 0-4; 0=no debris present, 4=debris present more than two thirds of area beneath lens.

    Measured: Baseline

  7. Lens Surface - Debris - Hema-copolymer and Etafilcon A

    Lens surface of debris for hema-copolymer and etafilcon A pair of lenses assessed at 1 week. (Each pair of lenses worn by the participant was assigned a single grade). Overall score measured by Grade 0-4; 0=no debris present, 4=debris present more than two thirds of area beneath lens.

    Measured: 1 week

  8. Lens Surface - Wettability - Hema-copolymer and Etafilcon A

    Lens surface of wettability for hema-copolymer and etafilcon A pair of lenses assessed at baseline. (Each pair of lenses worn by the participant was assigned a single grade). Overall score measured by Grade 0-4; 0=fully wetting lens surface, 4=presence of one or more non-wetting areas.

    Measured: Baseline

  9. Lens Surface - Wettability - Hema-copolymer and Etafilcon A

    Lens surface of wettability for hema-copolymer and etafilcon A pair lenses assessed at 1 week. (Each pair of lenses worn by the participant was assigned a single grade). Overall score measured by Grade 0-4; 0=fully wetting lens surface, 4=presence of one or more non-wetting areas.

    Measured: 1 week

  10. Lens Fit - Horizontal Centration - Hema-copolymer and Etafilcon A

    Lens fit of horizontal centration for hema-copolymer and etafilcon A lenses assessed at baseline. Overall score measured by extremely nasal, slightly nasal, optimum, slightly temporal, extremely temporal

    Measured: Baseline

  11. Lens Fit - Horizontal Centration - Hema-copolymer and Etafilcon A

    Lens fit of horizontal centration for hema-copolymer and etafilcon A lenses assessed at 1 week. Overall score measured by extremely nasal, slightly nasal, optimum, slightly temporal, extremely temporal

    Measured: 1 week

  12. Lens Fit - Vertical Centration - Hema-copolymer and Etafilcon A

    Lens fit of vertical centration for hema-copolymer and etafilcon A lenses assessed at baseline. Overall score measured by extremely inferior, slightly inferior, optimum, slightly superior, extremely superior

    Measured: Baseline

  13. Lens Fit - Vertical Centration - Hema-copolymer and Etafilcon A

    Lens fit of vertical centration for hema-copolymer and etafilcon A lenses assessed at 1 week. Overall score measured by extremely inferior, slightly inferior, optimum, slightly superior, extremely superior

    Measured: 1 week

  14. Lens Fit - Corneal Centration - Hema-copolymer and Etafilcon A

    Lens fit of corneal centration for hema-copolymer and etafilcon A lenses assessed at baseline. Overall score measured by extremely inadequate, slightly inadequate, optimum, slightly excessive, extremely excessive

    Measured: Baseline

  15. Lens Fit - Corneal Centration - Hema-copolymer and Etafilcon A

    Lens fit of corneal centration for hema-copolymer and etafilcon A lenses assessed at 1 week. Overall score measured by extremely inadequate, slightly inadequate, optimum, slightly excessive, extremely excessive

    Measured: 1 week

  16. Lens Fit - Lens Movement - Hema-copolymer and Etafilcon A

    Lens fit of lens movement for hema-copolymer and etafilcon A lenses assessed at baseline. Overall score measured by extremely inadequate, slightly inadequate, optimum, slightly excessive, extremely excessive

    Measured: Baseline

  17. Lens Fit - Lens Movement - Hema-copolymer and Etafilcon A

    Lens fit of lens movement for hema-copolymer and etafilcon A lenses assessed at 1 week. Overall score measured by extremely inadequate, slightly inadequate, optimum, slightly excessive, extremely excessive

    Measured: 1 week

  18. Comfort Preference (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of comfort preference for hema-copolymer and etafilcon A assessed at baseline. Scale 0-100, 0=causes pain, 100= excellent.

    Measured: Baseline

  19. Comfort Preference (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of comfort preference after insertion and before removal for hema-copolymer and etafilcon A lenses is assessed at 1 week. Scale 0-100, 0=causes pain, 100= excellent.

    Measured: 1 week

  20. Vision Preference (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of vision preference for hema-copolymer and etafilcon A lenses assessed at baseline. Scale 0-100, 0=unacceptable, 100= excellent.

    Measured: Baseline

  21. Vision Preference (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of vision preference for hema-copolymer and etafilcon A lenses assessed at 1 week. Scale 0-100, 0=unacceptable, 100= excellent.

    Measured: 1 week

  22. Peripheral Blur (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of peripheral blur is assessed at baseline. Scale 0-100, 0=unacceptable, 100= excellent.

    Measured: Baseline

  23. Peripheral Blur (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of peripheral blur assessed at 1 week. Scale 0-100, 0=unacceptable, 100= excellent.

    Measured: 1 week

  24. Ocular Redness (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of ocular redness for hema-copolymer and etafilcon A lenses lenses assessed at 1 week. Scale 0-100, 0=extremely poor, 100= excellent.

    Measured: 1 week

  25. Handling (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of handling (ease of insertion and ease of removal) for hema-copolymer and etafilcon A lenses assessed at 1 week. Scale 0-100, 0=unmanageable, 100= excellent.

    Measured: 1 week

  26. Attractiveness (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of attractiveness for hema-copolymer and etafilcon A lenses assessed at baseline. Scale 0-100, 0=extremely poor, 100= excellent.

    Measured: Baseline

  27. Attractiveness (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective Assessment of attractiveness for hema-copolymer and etafilcon A lenses assessed at 1 week. Scale 0-100, 0=extremely poor, 100= excellent.

    Measured: 1 week

  28. Overall Impression (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective assessment of the overall impression for hema-copolymer and etafilcon A lenses assessed at baseline. Scale 0-100, 0=extremely poor, 100= excellent.

    Measured: Baseline

  29. Overall Impression (Subjective Assessment) - Hema-copolymer and Etafilcon A

    Subjective assessment of the overall impression for hema-copolymer and etafilcon A lenses assessed at 1 week. Scale 0-100, 0=extremely poor, 100= excellent.

    Measured: 1 week

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted18 Oct 2013 (estimated)
  2. Study startApr 2014
  3. Primary completionMay 2014
  4. Study completionOct 2014

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Masking
Single blind
Who is masked
Participant
Enrolment
30 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01965262
Sponsor's protocol ID
CV13-545
Status last verified
Feb 2017
First submitted
11 Oct 2013
Last updated
16 Mar 2017
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing Hema-copolymer Lens and etafilcon A Lens in people with Myopia, run by CooperVision, Inc. at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 8 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 16 Mar 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 18 Apr 2016). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Visual Acuity - Hema-copolymer and Etafilcon A (time frame: Baseline). The study lists 29 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Mar 2017.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in