Effect of Different Plasma Insulin Levels on the Accuracy of Continuous Subcutaneous Glucose Monitoring

A study testing High insulin eu-hypoglycemic clamp and Low insulin eu-hypoglycemic clamp in people with Type 1 Diabetes and Continuous Glucose Monitoring, run by Fundación para la Investigación del Hospital Clínico de Valencia at 1 site in Spain.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Achieving near-normoglycaemia has been established as the main objective for most patients with diabetes. However, it is well known that intensification of treatment is associated with an increase in the frequency of hypoglycemia, especially in the context of insulin therapy. Read more about this study

Who is it for?
Ages 18 to 60 years
What is being tested?
High insulin eu-hypoglycemic clamp and Low insulin eu-hypoglycemic clamp
Type of treatment
Injection or infusion
Could I receive a placebo?
No, everyone receives an active treatment
How many people
14 participants
Where
1 siteSpain
Study dates
Oct 2011to Oct 2012
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 60 years
  • All sexes
  • No healthy volunteers
  • Adult

Who may be able to join

  • Subjects with type 1 diabetes mellitus meeting all of the following criteria will be considered for admission to the study:
  • Aged between 18 and 60 years
  • Under CSII or MDI treatment for at least six months before Visit 1
  • Body mass index of between 18 and 30 kg/m2
  • HbA1c 6.0-8.5% at Visit 1
  • Normal laboratory values, ECG, and vital signs unless the investigator considered an abnormality to be clinically irrelevant
  • Women postmenopausal or using contraception judged by the investigator to be adequate (e.g., oral contraceptives, intra-uterine device or surgical treatment)

Who cannot join

  • Subjects meeting any of the following criteria will not be included in the study:
  • Pregnancy and lactation
  • History of hypersensitivity to the study medications or to drugs with similar chemical structures
  • Confirmed hypoglycaemia unawareness
  • Progressive fatal diseases
  • History of drug or alcohol abuse
Show all 15Show fewer
  • History of positive HIV or hepatitis B or C test
  • Impaired hepatic function, as shown by, but not limited to, SGPT or SGOT of more than twice the upper limit of the normal range at visit 1
  • Impaired renal function, as shown by, but not limited to, serum creatinine > 1.5 mg/dL at visit 1
  • Clinically relevant microvascular, cardiovascular, hepatic, neurologic, endocrine or other major systemic diseases other than T1DM which could hinder implementation of the clinical study protocol or interpretation of the study results
  • Pre-planned surgery during the study
  • Blood donation of more than 500 ml during the past three months for men, or during the past six months for women
  • Mental condition rendering the subject unable to understand the nature, scope and possible consequences of the study
  • Subject unlikely to comply with clinical study protocol, e.g., uncooperative attitude, inability to return for follow-up visits, or poor likelihood of completing the study
  • Receipt of an experimental drug or use of an experimental device during the past 30 days.

Trial locations and contacts

1 site in 1 country.

Spain1 site
  • Servicio de Endocrinología y Nutrición, Hospital Clínico Universitario de Valencia

    Valencia, 46010

Other trials in Spain

About this study

Official title: New Strategies for Automated Glycaemic Control: the Issue of Continuous Glucose Monitoring Accuracy Under Hypoglycaemic Conditions

Achieving near-normoglycaemia has been established as the main objective for most patients with diabetes. However, it is well known that intensification of treatment is associated with an increase in the frequency of hypoglycemia, especially in the context of insulin therapy. The burden of hypoglycemia in terms of psychological implications, morbidity and even mortality, explains why it has been defined as the main limiting factor to achievement of good metabolic control.

Continuous subcutaneous glucose monitoring (CGM) devices have been claimed to be useful in hypoglycemia detection/prevention, allowing theoretically for safer intensification of therapy in diabetic patients. However, accuracy of CGM devices, especially in the hypoglycemic range, raises some concerns. In fact, commercially available CGM devices estimate plasma glucose from measurements in the interstitial fluid and not in plasma. However, the relationship between plasma and interstitial glucose is not fully understood, especially under dynamic conditions, and this may explain the poor CGM performance during rapid changes in blood glucose and hypoglycemia.

Read more

In this project, the relationship between plasma and interstitial glucose will be evaluated under conditions of normal glucose concentrations and hypoglycemia. Experiments will be performed to assess the role, if any, of different plasma insulin concentrations on the accuracy of CGM.

All the information obtained may be relevant to the improvement of the ability of CGM devices to detect hypoglycemia and hypoglycemic risk.

Being tested
High insulin eu-hypoglycemic clamp, Low insulin eu-hypoglycemic clamp
Keywords
Interstitial glucose dynamics, Accuracy of continuous glucose monitoring under hypoglycemia, Glucose clamp
Medical subject headings
Diabetes Mellitus, Type 1, Hypoglycemia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • High insulin eu-hypoglycemic clampOther

    Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.

  • Low insulin eu-hypoglycemic clampOther

    Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.

Study groups in NCT01714895
Study groupRoleWhat they receive
High plasma insulin-Low plasma insulinOtherSeven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a High insulin glucose clamp and on the second day a Low insulin glucose clamp (sequence 'AB')
Receives: Other: High insulin eu-hypoglycemic clamp; Other: Low insulin eu-hypoglycemic clamp
Low plasma insulin-High plasma insulinOtherSeven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a Low insulin glucose clamp and on the second day a High insulin glucose clamp (sequence 'BA')
Receives: Other: High insulin eu-hypoglycemic clamp; Other: Low insulin eu-hypoglycemic clamp

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Normalized current signal from the CGM devices

    CGM devices are electrochemical sensors. Currently available sensors are placed in the subcutaneous tissue where they react with the glucose of the interstitial fluid giving a current signal (in nano Amperes), which is proportional to glucose concentrations. The current signal is then filtered and transformed into plasma glucose estimations through a calibration algorithm built into the CGM device.

    In this study we will analyze the current signal from CGM obtained under two different experimental conditions:

    1. Eu-hypoglycemic clamp with High plasma insulin concentrations
    2. Eu-hypoglycemic clamp with relatively Low plasma insulin concentrations

    This will allow for establishing the role, if any, of different plasma insulin concentrations on the accuracy of CGM to identify plasma glucose falls resulting in hypoglycaemia.

    Since CGM sensors can exhibit different sensitivities to glucose concentrations, the current signal will be normalized before statistical analysis

    Measured: 330 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted26 Oct 2012 (estimated)
  2. Study startOct 2011
  3. Primary completionSep 2012
  4. Study completionOct 2012

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Basic science
Allocation
Randomised
Intervention model
Crossover
Masking
Double blind
Who is masked
Participant, Outcomes assessor
Enrolment
14 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01714895
Sponsor's protocol ID
FP7-PEOPLE-2009-IEF #252085-2
Other IDs
Spanish Ministry of Science
Status last verified
Oct 2012
First submitted
24 Oct 2012
Last updated
26 Oct 2012
Results posted
No
Data monitoring committee
No

Publications

  • Moscardo V, Bondia J, Ampudia-Blasco FJ, Fanelli CG, Lucidi P, Rossetti P. Plasma Insulin Levels and Hypoglycemia Affect Subcutaneous Interstitial Glucose Concentration. Diabetes Technol Ther. 2018 Apr;20(4):263-273. doi: 10.1089/dia.2017.0219. Epub 2018 Mar 12.

    PubMed 29638161

Questions about this trial

What is this clinical trial studying?

A study testing High insulin eu-hypoglycemic clamp and Low insulin eu-hypoglycemic clamp in people with Type 1 Diabetes and Continuous Glucose Monitoring, run by Fundación para la Investigación del Hospital Clínico de Valencia at 1 site in Spain.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 15 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 26 Oct 2012. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Spain. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Normalized current signal from the CGM devices (time frame: 330 minutes). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01714895, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.