Clinical Study of Wrist Blood Pressure Monitor in Conformance With the ANSI/AAMI SP10 Standard

A study testing Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117 and Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2 in people with Efficacy of Wrist BPM Measurement and Hypertension, run by Leo Wang at 1 site in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The clinical protocol of the clinical testing of this device: 1. Objective of the test: To verify the efficiency of device. 2. Test methods and procedures: Clinical test in two positions: Seated and Supine. 3. DUT: Transtek Wrist Blood Pressure Monitor, Model: TMB-1117. Cuff size: 13.5 - 21.5 cm. 4. Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
What is being tested?
Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117 and Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
85 participants
Where
1 siteChina
Study dates
May 2012to Jun 2012
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • male,female,

Who cannot join

  • below 18 years old

Trial locations and contacts

1 site in 1 country.

China1 site
  • Zhongshan City People's Hospital

    Zhongshan, Guangdong, 528403

Other trials in China

About this study

Official title: Verify Wrist BPM TMB-1117 to Comply With ANSI/AAMI SP10

The clinical protocol of the clinical testing of this device:

  1. Objective of the test: To verify the efficiency of device.
  2. Test methods and procedures: Clinical test in two positions: Seated and Supine.
  3. DUT: Transtek Wrist Blood Pressure Monitor, Model: TMB-1117. Cuff size: 13.5 - 21.5 cm.
  4. Comparison device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1 mmHg and range: 0-300 mmHg.
  5. Study endpoints: Comply with ANSI/AAMI SP10-2002 standard.
  6. Statistical methodology used: Description of statistical methods.
  7. Result: Meet the requirements of SP10.
Read the detailed description

Hospital Information

Hospital Information The data was collected by Zhongshan City People's Hospital Clinical Investigator Team at Zhongshan City People's Hospital, No.2 Sunwen Dong Road, Zhongshan, Guangdong 528403, P. R. China.

Investigator: Dr. Cao Yuedong, Leader; Li Qidong, Nurse A; Chen Wenqian, Nurse B.

Contact Dr. Cao Yuedong Tel: +86 760 88823818

Used Equipments DUT (Device Under Test): Transtek Wrist Blood Pressure Monitor, TMB-1117 Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1 mmHg and range: 0-300 mmHg.

Test Protocol

  1. Test Purpose: The aim of clinical test is to determine, for the purpose of design qualification, the overall performance of the system meeting the following requirement: For systolic and diastolic pressures, treated separately, the mean difference of the paired measurement of the test system and the comparison system shall be ± 5 mmHg or less, with standard deviation of 8 mmHg or less.
  2. Target Subject:
  3. The subject database shall contain at least 85 subjects. 2) The distribution of blood pressure of the subjects required to be: At least 10 percent below 100 mmHg systolic; at least 10 percent above 160 mmHg systolic; at least 10 percent below 60 mmHg diastolic; at least 10 percent above 100 mmHg diastolic; with the remainder distributed between these outer limits.
  4. Identification and precision of equipment used According to ANSI/AAMI SP10, the maximum measurement error of the non-automatic sphygmomanometers used in the comparing test shall not exceed 1.0mmHg at the temperature of the test. The manometer was used to verify its accuracy before test.

The clinical study was conducted according to the protocol described in the ANSI/AAMI SP10 with 85 subjects.

  1. Test procedures (auscultatory method, same-limb sequential measurement was chosen)
  2. Record the age, gender, and circumference of the left wrist of the subject.
  3. Let subject seated or supine calmly for five minutes before test.
  4. Measure heart rate in one minute by palpation on radial artery, record it.
  5. Two observers shall make simultaneous, blinded, blood pressure determination on the subject's left arm. Each nurse will write down the reading on a small piece of paper, make a tick if there is an auscultatory gap and then submit to the recorder.
  6. The two readings separately from two observers will be recorded, and if two observers agree that they find an auscultatory gap this time, it will also be recorded. The wrist cuff will be take off, after 90 seconds rest, TMB-1117 cuff will be taken on and tested, and the readings (systolic, diastolic, mean BP and Heart rate) be recorded.
  7. For each subject, repeat 3) and 5) to get 3 measurements.
  8. Repeat step 1) to 6) for every subject.
  9. Note:
  10. No motion and speaking are allowed during the measurement.
  11. The reference device's cuff release at the rate about 3 mmHg/s so to ensure an accurate reading of observer.
  12. The arm, wrist and the haear of subject to keep on the same height when measurement.
  13. Test environment: Temperature: 20±1 ℃; Relative humidity: 40~50%.
Being tested
Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117, Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2
Keywords
blood pressure
Medical subject headings
Hypertension

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117Device

    Also called DUT TMB-1117 measurement

    Systolic, Diastolic pressure.

  • Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2Device

    Also called Reference mercury sphygmomanometer measurement

    Systolic, Diastolic pressure

Study groups in NCT01662947
Study groupRoleWhat they receive
DUT ArmExperimentalDUT: Transtek Wrist Blood Pressure Monitor TMB-1117 Measurement: Blood Pressure Groups/Cohorts: DUT
Receives: Device: Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117
Reference ArmExperimentalReference Device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1mmHg and range: 0-300mmHg. Measurement: Blood Pressure Groups/Cohorts: Reference
Receives: Device: Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Verify the accuracy of measure functions of device

    Verify the accuracy of device if meet the requirements of ANSI/AAMI SP10 when it measure Systolic \& Diastolic blood pressure.

    Measured: 10 days

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 Aug 2012 (estimated)
  2. Study startMay 2012
  3. Primary completionJun 2012
  4. Study completionJun 2012

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Factorial
Masking
None (open label)
Enrolment
85 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01662947
Sponsor's protocol ID
Wrist BPM TMB-1117
Other IDs
BTS-TRANS12001
Acronym
WristBPM01
Status last verified
Aug 2012
First submitted
1 Aug 2012
Last updated
13 Aug 2012
Results posted
No
Data monitoring committee
No

Publications

  • Alpert BS. Validation of the Tiba Medical Ambulo 2400 ambulatory blood pressure monitor to the ISO Standard and BHS protocol. Blood Press Monit. 2010 Oct;15(5):275-7. doi: 10.1097/MBP.0b013e32833c8b39.

    PubMed 20559140

Questions about this trial

What is this clinical trial studying?

A study testing Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117 and Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2 in people with Efficacy of Wrist BPM Measurement and Hypertension, run by Leo Wang at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 13 Aug 2012. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Verify the accuracy of measure functions of device (time frame: 10 days). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01662947, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 13 Aug 2012.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in