Safety of Screening Procedures With Hand-held Metal Detectors Among Patients With Implanted Cardiac Rhythm Devices

A study testing Hand-held metal detector in people with Electromagnetic Interference, run by Deutsches Herzzentrum Muenchen at 2 sites in 2 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Patients with implanted pacemaker (PM) or cardioverter-defibrillator (ICD) systems are restrained in daily life by possible electromagnetic interference (EMI). Case reports suggest EMI between PM or ICD systems and hand-held metal detectors that are intensively used as part of security screening processes in e.g. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Electromagnetic Interference
What is being tested?
Hand-held metal detector
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
388 participants
Where
2 sitesin 2 countries
Study dates
Sep 2009to Dec 2010
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients presenting for routine pacemaker or cardioverter defibrillator control
  • appropriate PM/ICD function

Who cannot join

  • atrial or ventricular sensing abnormalities that could not be avoided by reprogramming of device parameters
  • low battery status of the device such as elective replacement indicator (ERI)
  • intrinsic heart rate > 120 beats per minute (bpm)

Trial locations and contacts

2 sites in 2 countries.

2 sites in 2 countries

Germany1 site
  • Deutsches Herzzentrum München

    München, 80636

Other trials in Germany

Greece1 site
  • Cardiology Department, Hospital "Henry Dunant"

    Athens

Other trials in Greece

About this study

Patients with implanted pacemaker (PM) or cardioverter-defibrillator (ICD) systems are restrained in daily life by possible electromagnetic interference (EMI). Case reports suggest EMI between PM or ICD systems and hand-held metal detectors that are intensively used as part of security screening processes in e.g. airport controls.

The objective was to determine the safety of screening procedures for ferrous materials with regard to possible in vivo EMI between hand-held metal detectors and PM and ICD systems.

Being tested
Hand-held metal detector
Keywords
electromagnetic interference, hand-held metal detector, pacemaker, cardioverter-defibrillator

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Hand-held metal detectorDevice

    2 different hand-held metal detectors: (1) PD 140 (CEIA S.p.A., Arezzo, Italy) and (2) MH 5 (Vallon GmbH, Eningen, Germany)

Study groups in NCT01369706
Study groupRoleWhat they receive
Hand-held metal detectorOtherExposure to two hand-held metal detectors
Receives: Device: Hand-held metal detector

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Electromagnetic Interference

    inhibition of the pacemaker, loss of capture, inappropriate mode switch, ventricular oversensing, power-on-reset, device reprogramming or loss of function

    Measured: time during exposure to hand-held metal detector (2x 30 sec)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted9 Jun 2011 (estimated)
  2. Study startSep 2009
  3. Primary completionDec 2010
  4. Study completionDec 2010

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
388 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01369706
Sponsor's protocol ID
GER-EP-007
Status last verified
Apr 2017
First submitted
6 Jun 2011
Last updated
12 Apr 2017
Data monitoring committee
No

Publications

  • Jilek C, Tzeis S, Vrazic H, Semmler V, Andrikopoulos G, Reents T, Fichtner S, Ammar S, Rassias I, Theodorakis G, Weber S, Hessling G, Deisenhofer I, Kolb C. Safety of screening procedures with hand-held metal detectors among patients with implanted cardiac rhythm devices: a cross-sectional analysis. Ann Intern Med. 2011 Nov 1;155(9):587-92. doi: 10.7326/0003-4819-155-9-201111010-00005.

    PubMed 22041947

Questions about this trial

What is this clinical trial studying?

A study testing Hand-held metal detector in people with Electromagnetic Interference, run by Deutsches Herzzentrum Muenchen at 2 sites in 2 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 12 Apr 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in Germany, Greece. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 30 Nov 2015). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Electromagnetic Interference (time frame: time during exposure to hand-held metal detector (2x 30 sec)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01369706, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 12 Apr 2017.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in