Evaluation of 3 Intraocular Lenses Following Lens Extraction

A Phase 4 study testing Crystalens AO, ReSTOR and Tecnis Multifocal IOL in people with Cataract, run by Bausch & Lomb Incorporated at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The objective of this study is to evaluate the efficacy and subject satisfaction with three different FDA approved lenses for adults over 40 years of age who desire a reduction in spectacle dependence. Read more about this study

Who is it for?
Ages 40 years and older
Conditions the study is for
Cataract
What is being tested?
Crystalens AO, ReSTOR and Tecnis Multifocal IOL
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
78 participants
Where
1 siteUnited States
Study dates
Jun 2010to Sep 2012
Phase
Phase 4
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 40 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subjects must have a clinically documented diagnosis of age-related bilateral cataracts that are considered amenable to treatment with standard phacoemulsification cataract extraction.
  • Subjects must require a spherical lens power from 10.00 D to 33.00 D
  • Subjects must have the potential for corrected distance visual acuity (CDVA) of 20/32 or better in both eyes, as documented by a Potential Acuity Meter

Who cannot join

  • Subjects with conditions with increased risk of zonular rupture, such as pseudoexfoliation syndrome. Zonular rupture during cataract surgery may affect postoperative centration, tilt, and stability of the lens.
  • Subjects who have any active inflammation or edema (swelling) of the cornea, including but not limited to the following: keratitis, keratoconjunctivitis, and keratouveitis.
  • Subjects with visually significant diabetic retinopathy (proliferative or non-proliferative) which reduces potential acuity to 20/32 or worse.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Bausch & Lomb, Inc

    Aliso Viejo, California, 92656

Other trials in United States

About this study

Official title: A Three Arm Prospective Clinical Evaluation of Three FDA Approved Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction.

The objective of this study is to evaluate the efficacy and subject satisfaction with three different FDA approved lenses for adults over 40 years of age who desire a reduction in spectacle dependence.

Conditions
Cataract
Being tested
Crystalens AO, ReSTOR, Tecnis Multifocal IOL
Keywords
Intra Ocular Lens, Ocular surgery, Phacoemulsification
Medical subject headings
Cataract

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Crystalens AODevice

    Crystalens AO surgically implanted bilaterally. Study observation up to 180 days.

  • ReSTORDevice

    ReSTOR surgically implanted bilaterally. Study observation up to 180 days.

  • Tecnis Multifocal IOLDevice

    Tecnis surgically implanted bilaterally. Study observation up to 180 days.

Study groups in NCT01122576
Study groupRoleWhat they receive
Crystalens AOExperimentalEligible subjects to undergo small incision cataract surgery and were implanted with the Crystalens AO bilaterally.
Receives: Device: Crystalens AO
ReSTORActive comparatorEligible subjects to undergo small incision cataract surgery and were implanted with the ReSTOR IOL bilaterally.
Receives: Device: ReSTOR
Tecnis Multifocal IOLActive comparatorEligible subjects to undergo small incision cataract surgery and were implanted with the Tecnis Multifocal IOL bilaterally.
Receives: Device: Tecnis Multifocal IOL

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mesopic Monocular Contrast Sensitivity Without Glare 1.5 and 3 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 3 (2-3 months)

  2. Mesopic Monocular Contrast Sensitivity Without Glare 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 3 (2-3 months)

  3. Mesopic Monocular Contrast Sensitivity Without Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 3 (2-3 months)

  4. Mesopic Monocular Contrast Sensitivity Without Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 3 (2-3 months)

  5. Mesopic Monocular Contrast Sensitivity Without Glare 1.5 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 4 ( 4-6 months)

  6. Mesopic Monocular Contrast Sensitivity Without Glare 3 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 4 ( 4-6 months)

  7. Mesopic Monocular Contrast Sensitivity Without Glare 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 4 ( 4-6 months)

  8. Mesopic Monocular Contrast Sensitivity Without Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 4 ( 4-6 months)

  9. Mesopic Monocular Contrast Sensitivity Without Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Postoperative visit 4 ( 4-6 months)

Secondary outcomes (35)
  1. Binocular Mesopic Contrast Sensitivity Without Glare 1.5 & 3 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  2. Binocular Mesopic Contrast Sensitivity Without Glare 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  3. Binocular Mesopic Contrast Sensitivity Without Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  4. Binocular Mesopic Contrast Sensitivity Without Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  5. Binocular Mesopic Contrast Sensitivity Without Glare 1.5, 3, 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  6. Binocular Mesopic Contrast Sensitivity Without Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  7. Binocular Mesopic Contrast Sensitivity Without Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  8. Mesopic Monocular Contrast Sensitivity With Glare 1.5 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  9. Mesopic Monocular Contrast Sensitivity With Glare 3 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  10. Mesopic Monocular Contrast Sensitivity With Glare 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  11. Mesopic Monocular Contrast Sensitivity With Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  12. Mesopic Monocular Contrast Sensitivity With Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  13. Mesopic Monocular Contrast Sensitivity With Glare 1.5 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  14. Mesopic Monocular Contrast Sensitivity With Glare 3 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  15. Mesopic Monocular Contrast Sensitivity With Glare 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  16. Mesopic Monocular Contrast Sensitivity With Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  17. Mesopic Monocular Contrast Sensitivity With Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  18. Mesopic Binocular Contrast Sensitivity With Glare 1.5, 3 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  19. Mesopic Binocular Contrast Sensitivity With Glare 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  20. Mesopic Binocular Contrast Sensitivity With Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  21. Mesopic Binocular Contrast Sensitivity With Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 3 (2-3 months)

  22. Mesopic Binocular Contrast Sensitivity With Glare 1.5, 3 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  23. Mesopic Binocular Contrast Sensitivity With Glare 6 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  24. Mesopic Binocular Contrast Sensitivity With Glare 12 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  25. Mesopic Binocular Contrast Sensitivity With Glare 18 Cycles/Degree

    Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity

    Measured: Visit 4 (4-6 months)

  26. Optical Scatter

    Objective scatter index (OSI) was assessed using the Optical Quality Analysis System(OQAS) and summarized as a continuous variable. For this assessment, an OSI score of 1 or less represents good quality vision, 1-2 indicates some degradation but otherwise normal vision, 2-3 significant light scatter possibly requiring treatment, and scores over 4 severely compromised vision requiring intervention.

    Measured: Visit 3 (2-3 months)

  27. Optical Scatter

    Objective scatter index (OSI) was assessed using the Optical Quality Analysis System(OQAS) and summarized as a continuous variable. For this assessment, an OSI score of 1 or less represents good quality vision, 1-2 indicates some degradation but otherwise normal vision, 2-3 significant light scatter possibly requiring treatment, and scores over 4 severely compromised vision requiring intervention.

    Measured: Visit 4 (4-6 months)

  28. Monocular Halo and Starburst

    Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.

    Measured: Visit 3 (2-3 months)

  29. Monocular Halo and Starburst

    Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.

    Measured: Visit 4 (4-6 months)

  30. Binocular Halo and Starburst

    Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.

    Measured: Visit 3 (2-3 months)

  31. Binocular Halo and Starburst

    Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.

    Measured: Visit 4 (4-6 months)

  32. Manifest Refraction

    Manifest refraction spherical equivalent (MRSE) was calculated as the value of the sphere plus one-half of the value of the cylinder.

    Measured: Visit 3 (2-3 Months)

  33. Manifest Refraction

    Manifest refraction spherical equivalent (MRSE) was calculated as the value of the sphere plus one-half of the value of the cylinder.

    Measured: Visit 4 (4-6 Months)

  34. Uncorrected Distance Visual Acuity

    High Contrast Uncorrected Distance (UCDVA) measured at 32 inches, Intermediate (UCIVA) at 32 inches, and Near Visual Acuity (UCNVA) at 16 inches

    Measured: Visit 3 (2-3 months) Visit 4 (4-6 months)

  35. Best-corrected Distance Visual Acuity

    High Contrast Distance-Corrected Distance,(BCDVA), Intermediate (DCIVA), and Near Visual Acuity (DCNVA)

    Measured: Visit 3 (2-3 months) Visit 4 (4-6 months)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 May 2010 (estimated)
  2. Study startJun 2010
  3. Primary completionJun 2012
  4. Study completionSep 2012

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase 4
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
Double blind
Who is masked
Participant, Outcomes assessor
Enrolment
78 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01122576
Sponsor's protocol ID
651
Status last verified
Feb 2014
First submitted
11 May 2010
Last updated
28 Mar 2014
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A Phase 4 study testing Crystalens AO, ReSTOR and Tecnis Multifocal IOL in people with Cataract, run by Bausch & Lomb Incorporated at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 28 Mar 2014. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

What does this phase mean?

Phase 4 studies take place after a treatment is approved, to learn more about its long-term benefits and risks in everyday use.

Not every study follows the same path, and some studies (observational studies, device studies) have no phase.

Are results available for this study?

Yes. Results are posted in the registry (first posted 28 Mar 2014). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mesopic Monocular Contrast Sensitivity Without Glare 1.5 and 3 Cycles/Degree (time frame: Postoperative visit 3 (2-3 months)). The study lists 9 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 28 Mar 2014.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in