Cardiac CatheterizationNCT00849940

Validation of Brain Oxygenation Monitor on Pediatric Patients

A study testing CAS NIRS FORE-SIGHT oximeter in people with Cardiac Catheterization, run by Duke University at 1 site in United States.

Terminated

Stopped early. The study ended before it was planned to and will not resume.

Official record
Search all trials

Trial summary

The aim of this study is calibrate (adjust and tune) the CAS FORE-SIGHT Near-Infrared Spectroscopy (NIRS) monitor when used to measure the tissue oxygen saturation of internal organs (StO2). This is a measure of the amount of oxygen carried by the blood within the internal organs. Read more about this study

Who is it for?
Ages Up to 17 years
Conditions the study is for
Cardiac Catheterization
What is being tested?
CAS NIRS FORE-SIGHT oximeter
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
72 participants
Where
1 siteUnited States
Study dates
Feb 2009to Nov 2014
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Up to 17 years
  • All sexes
  • No healthy volunteers
  • Child

Who may be able to join

  • ≤ 18 years old
  • ≥ 2.5 kg and ≤ 40 kg weight
  • Undergoing cardiac catheterization at Duke pediatric cardiac catheterization laboratory

Who cannot join

  • Known structural neurologic or craniofacial disease
  • Arterial catheter placement not part of planned catheterization procedure
  • Anemia (hemoglobin < 10 g/dl)

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Duke University Health System

    Durham, North Carolina, 27710

Other trials in United States

About this study

Official title: Validation of the CAS Medical System, Inc. FORE-SIGHT Near-Infrared Spectroscopy (NIRS) Monitor in Pediatric Subjects for Viscerosomatic Applications

The aim of this study is calibrate (adjust and tune) the CAS FORE-SIGHT Near-Infrared Spectroscopy (NIRS) monitor when used to measure the tissue oxygen saturation of internal organs (StO2). This is a measure of the amount of oxygen carried by the blood within the internal organs. In addition the study will assess the degree of similarities between StO2 and mean mixed venous oxygen saturation - a measure of the amount of oxygen carried in the blood returning to the heart.

Read the detailed description

NIRS cerebral oximeters are FDA-approved devices used to measure oxygen saturation within the brain, in a similar manner to pulse oximeters that measure the oxygen saturation in the finger tip. The sensor pads are placed on the surface of the forehead and shine near-infrared light through the skull and brain tissue from which the brain tissue oxygen saturation is estimated. The same principles can be applied when the sensor pads are placed over the internal organs of the abdomen, for example, the liver. Currently the only way to accurately measure the oxygen saturation of internal organs is by the invasive placement of intravenous lines into the blood vessels of that organ. This study will determine if the NIRS sensors can reliably estimate the tissue oxygen saturation non-invasively by placing the pad over the skin of the abdomen.

The study will be conducted in pediatric patients who are undergoing cardiac catheterization, a procedure in which invasive lines are placed in order to get information about the heart. The procedure is always conducted under general anesthesia. During the cardiac catheterization procedure blood samples are routinely taken for oxygen saturation analysis. In the study two oximeter sensor pads will be placed on the forehead (one on each side) and two further oximeter sensor pads will be placed on the abdominal wall. The oxygen saturation values from all oximeter sensors will be recorded continuously throughout the cardiac catheterization procedure and will be compared to the oxygen saturation values from the blood samples. In addition to the routine blood samples taken as part of the cardiac catheterization, one blood sample will be taken when the invasive line is within the right hepatic (liver) vein.

The information from this study will determine how well the oximeter sensors estimate the oxygen saturation of both the internal organs (StO2) and the blood returning to the heart (mean mixed venous oxygen saturation).

Being tested
CAS NIRS FORE-SIGHT oximeter
Keywords
Spectroscopy, Near infrared, Oximetry, Hypoxemia, Tissue oxygenation, Pediatric
Medical subject headings
Hypoxia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • CAS NIRS FORE-SIGHT oximeterDevice

    Comparison of non-invasive tissue oxygen saturation with blood sample-derived (calculated) tissue oxygen saturation.

Study groups in NCT00849940
Study groupRoleWhat they receive
CAS NIRS FORE-SIGHT oximeterExperimentalPediatric patients presenting for cardiac catheterization.
Receives: Device: CAS NIRS FORE-SIGHT oximeter

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Accuracy of NIRS Sensor to Estimate Flank Tissue Oxygen Saturation

    The flank tissue oxygen saturation (%) is determined from simultaneous arterial and venous blood samples processed through a blood gas machine. The blood oxygen saturation of the samples are entered into an equation to yield the best estimate of flank tissue oxygen saturation. This value is then compared to the NIRS oxygen saturation (%) displayed on the monitor. Accuracy is used to describe how close the NIRS oxygen saturation is to the flank tissue oxygen saturation. It can be expressed in terms of shots on target: bias (%) = how close are the shots to the bulls eye and precision (%) is how close are the shots to each other.

    oxygen saturation when measured displayed NIRS value of the tissue sensor placed over the flank to a reference CO-oximetry model, reported as bias and precision. The model is weighted as 30:70 arterial: central venous oxygen saturation when measured by blood gas co-oximetry.

    Measured: Data collected from individual participants over 4 hour timeframe.

  2. Accuracy of NIRS Sensor to Estimate Intestine Tissue Oxygen Saturation

    The intestine tissue oxygen saturation (%) is determined from simultaneous arterial and venous blood samples processed through a blood gas machine. The blood oxygen saturation of the samples are entered into an equation to yield the best estimate of intestine tissue oxygen saturation. This value is then compared to the NIRS oxygen saturation (%) displayed on the monitor. Accuracy is used to describe how close the NIRS oxygen saturation is to the intestine tissue oxygen saturation. It can be expressed in terms of shots on target: bias (%) = how close are the shots to the bulls eye and precision (%) is how close are the shots to each other.

    Measured: Data collected from individual participants over 4 hour timeframe.

  3. Accuracy of NIRS Sensor to Estimate Hepatic Tissue Oxygen Saturation

    The hepatic tissue oxygen saturation (%) is determined from simultaneous arterial and venous blood samples processed through a blood gas machine. The blood oxygen saturation of the samples are entered into an equation to yield the best estimate of hepatic tissue oxygen saturation. This value is then compared to the NIRS oxygen saturation (%) displayed on the monitor. Accuracy is used to describe how close the NIRS oxygen saturation is to the hepatic tissue oxygen saturation. It can be expressed in terms of shots on target: bias (%) = how close are the shots to the bulls eye and precision (%) is how close are the shots to each other.

    Measured: Data collected from individual participants over 4 hour timeframe.

Secondary outcomes (2)
  1. Correlation Between Somatic StO2 and Cerebral SctO2 Oxygen Saturation

    Measured: Data collected from individual participants over 4 hour timeframe

  2. Correlation Between NIRS Derived Estimate of Hemoglobin Concentration and Measured Arterial Blood Hemoglobin Concentration.

    Measured: Data collected from individual participants over 4 hour timeframe

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted24 Feb 2009 (estimated)
  2. Study startFeb 2009
  3. Primary completionNov 2014
  4. Study completionNov 2014

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
72 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00849940
Sponsor's protocol ID
Pro00009391
Status last verified
Feb 2016
First submitted
22 Jan 2009
Last updated
9 Feb 2017
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing CAS NIRS FORE-SIGHT oximeter in people with Cardiac Catheterization, run by Duke University at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: stopped early, as of the sponsor's last update on 9 Feb 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 28 Mar 2016). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Accuracy of NIRS Sensor to Estimate Flank Tissue Oxygen Saturation (time frame: Data collected from individual participants over 4 hour timeframe.). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Feb 2017.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in