CataractPhase 4NCT00459303

Comparison of Functional Vision Provided by AMO Tecnis Z9000 and Alcon SA60AT Acrysof

A Phase 4 study testing aspherical intraocular lens and spherical intraocular lens in people with Cataract, run by National Taiwan University Hospital at 1 site in Taiwan.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study is to compare intraindividually the functional vision provided by two different posterior chamber intraocular lenses: AMO Tecnis Z9000 and Alcon SA60AT Acrysoft. To see if the aspherical intraocular lenses provide better functional vision than traditional spherical intraocular lenses. Read more about this study

Who is it for?
Ages 45 to 80 yearsHealthy volunteers can take part
Conditions the study is for
Cataract
What is being tested?
aspherical intraocular lens and spherical intraocular lens
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
20 participants
Where
1 siteTaiwan
Study dates
Oct 2005to Dec 2007
Phase
Phase 4
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 45 to 80 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Bilateral cataracts, eligible for phacoemulsification with primary implantation of a posterior chamber IOL
  • Willing and able to comply with scheduled visits and other study procedures

Who cannot join

  • Preoperative ocular pathology potentially affect visual acuity. EX: diabetic retinopathy, macular degeneration, corneal opacity, glaucoma, visual field defect...etc.
  • Previous ocular surgery. Ex: refractive surgery, vitreoretinal surgery...etc.
  • Patients who cannot cooperative with the study procedures

Trial locations and contacts

1 site in 1 country.

Taiwan1 site
  • National Taiwan University Hospital

    Taipei

Other trials in Taiwan

About this study

Official title: Intraidivisual Comparison of Functional Vision Provided by AMO Tecnis Z9000 and ALcon SA60AT Acrysof Posterior Chamber Intraocular Lenses

This study is to compare intraindividually the functional vision provided by two different posterior chamber intraocular lenses: AMO Tecnis Z9000 and Alcon SA60AT Acrysoft. To see if the aspherical intraocular lenses provide better functional vision than traditional spherical intraocular lenses.

Read the detailed description

In this prospective study, 20 patients with bilateral catarct presenting for cataract surgery are randomly assigned to recieve an aspherical intraocular lens(AMO Tecnis Z9000) in one eye and a spherical intraocular lens(Alcon SA60AT Acrysof)in the other. All surgeries are performed by one experienced surgeon(Fung-Rong Hu,MD)with clear cornea small incision phacoemulsification and in-the-bag posterior chamber intraocular implantation.

The patients are followed for 3 months, and postoperative contrast sensitivity, contrast acuity, corneal \& total ocular high-ordered aberrations are measured and compared intraidividually between two different IOL groups.

Outcome Measurement:

  1. Contrast sensitivity testing: measured with spectacle correction for the target distance of three meters using wall-mounted FACT sine-wave grating chart with nine levels of contrast.The graphic contrast sensitivity is recorded as a log functional units on the y-axis and displayed for five spatial frequency targets (1.5, 3, 6, 12, 18 cycles per degree (cpd))on x- axis.
  2. Contrast acuity testing: measured with spectacle correction for the target distance of three meters using logMAR letter chart(Precision Vision®)represented on a wall-mounted illuminator cabinet.Two types of contrast charts are used, high contrast(Cat.No.2103 SLOAN translucent chart)and low contrast(Cat.No.2132 10% SLOAN translucent chart),and are tested under both photopic and mesopic conditions.
  3. Pupil diameters: measured by Colvard infrared pupillometer (Oasis. Medical, Glendora, CA, USA )under both photopic and mesopic luminance levels.
  4. Corneal aberration analysis: performed with TMS-4 (Tomey, Japan).The corneal HOAs are described with Zernike polynomials of 3rd-to 5th- order root-mean-square(RMS)of central 6mm diameter using VOLPro 6.89 software(Fa. Sarver and Associates, Carbondale. Ill, USA).
  5. Total ocular aberration analysis: performed with a Hartmann-Shack aberrometer(Zywave, Bausch \& Lomb Inc., Rochester, New. York)under maximal mydriasis with Mydrin-P(phenylephrine hydrochloride 0.5% and tropicamide 0.5%, Santen).The wavefront errors were described using Zernike polynomials RMS for total HOA at pupil diameter of 5 mm and 6 mm, primary spherical (Z 4,0) and 3rd-to 5th-order aberration at pupil diameter of 6 mm.

Data Analysis:

  • Mann-Whitney U matched-paired test was used with STATA software
  • P values of 0.05 or less were considered statistically significant
Conditions
Cataract
Being tested
aspherical intraocular lens, spherical intraocular lens
Keywords
cataract, aspherical intraocular lenses, functional vision
Medical subject headings
Cataract

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • aspherical intraocular lensDevice

    Also called AMO Tecnis Z9000

    implantation of aspherical intraocular lenses(AMO Tecnis Z9000)during cataract suregry (small cornea incision phacoemulsification, deliver with Monarch II device, in the bag implantation)

  • spherical intraocular lensDevice

    Also called Alcon SA60AT Acrysof

    implantation of spherical intraocular lenses(Alcon SA60AT Acrysof)during cataract suregry (small cornea incision phacoemulsification, deliver with Monarch II device, in the bag implantation)

Study groups in NCT00459303
Study groupRoleWhat they receive
intraocular lensActive comparatorpatients with bilateral clinical significant cataract reisiceved cataract surgeries and recieved spherial intraocuar lens(SA60AT, Alcon) in one eye and aspherical intraocular lens(Tecnis Z9000, AMO)in the other respectively.
Receives: Device: aspherical intraocular lens; Device: spherical intraocular lens

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Best Corrected logMAR Contrast Acuity at Photopic/Mesopic Condition

    Contrast acuity testing was measured with spectacle correction for the target distance of three meters using a logMAR letter chart (Precision Vision®) represented on a wall-mounted illuminator cabinet. Two types of contrast charts were used, high contrast (Cat.No.2103 SLOAN translucent chart) and low contrast (Cat.No.2132 10% SLOAN translucent chart), and these were tested under both photopic (250 Lux) and mesopic (0.5 Lux) conditions. The contrast acuity tested ranges from 20/160 to 20/20(from worse to best), which equals LogMAR(Logarithm of the Minimum Angle of Resolution)0.9 to -0.3. All of the visual acuity and functional vision testing examinations mentioned above were performed by a single ophthalmologist. The outcomes were recorded as average of three post-operative measurments.

    Measured: average data of post-op 3rd, 6th, 12th week measurements

  2. Best Corrected Contrast Sensitivity in Photopic Condition

    Contrast sensitivity testing was measured with spectacle correction for the target distance of three meters using a wall-mounted FACT sine-wave grating chart with nine levels of contrast (Stereo Optical Inc.)and five spatial frequency targets (1.5, 3, 6, 12, 18 cycles per degree (cpd)) at 250 lux. The last correct grating seen for each spatial frequency is recorded and translated by the "EYEVIEW™ Functional Analysis Software" into a log contrast sensitivity unit. The outcomes were recorded as average of the three post-operative measurements.

    ( physiological range of contrast sensitivity: 1.5 cpd : 25~82.5 ; 3 cpd: 30~150 ; 6 cpd: 65 ~ 200 ; 12 cpd: 20 ~130 ; 18 cpd: 6.5 ~ 65 )

    Measured: average data of post-operative 3rd week, 6th week, 12th week measurements

Secondary outcomes (2)
  1. Corneal High-order Aberrations

    Corneal topography was performed with a TMS-4 corneal tomographer (Tomey, Japan). We used the 31-rings placido-based system that covers 10.9 mm of corneal diameter which is sufficient for the study of aberrations up to the fifth order for 6 mm diameter. Corneal HOAs were described with Zernike polynomials of 3rd- to 5th-order root-mean-square (RMS) of central 6mm diameter using VOLPro 6.89 software (Fa. Sarver and Associates, Carbondale. Ill, USA).

    Measured: pre-op & averate data of post-op 3rd, 6th, 12th week measurements

  2. Total Ocular High-order Aberrations

    A Hartmann-Shack aberrometer (Zywave, Bausch \& Lomb Inc., Rochester, New York ) was used for measurement of HOAs of the whole eye. The measurements were done under maximal mydriasis with Mydrin-P (phenylephrine hydrochloride 0.5% and tropicamide 0.5%, Santen). The wavefront errors were described using the RMS of Zernike polynomials for total HOA at pupil diameters of 5 mm and 6 mm, primary spherical (Z 4,0) and 3rd-to 5th-order aberration at the pupil diameter of 6 mm.

    Measured: average data of post-op 3rd, 6th, 12th week measurements

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 Apr 2007 (estimated)
  2. Study startOct 2005
  3. Primary completionSep 2007
  4. Study completionDec 2007

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase 4
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
Double blind
Who is masked
Participant, Outcomes assessor
Enrolment
20 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00459303
Sponsor's protocol ID
05012006
Status last verified
Dec 2013
First submitted
10 Apr 2007
Last updated
17 Feb 2014
Data monitoring committee
Yes

Questions about this trial

What is this clinical trial studying?

A Phase 4 study testing aspherical intraocular lens and spherical intraocular lens in people with Cataract, run by National Taiwan University Hospital at 1 site in Taiwan.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 17 Feb 2014. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Taiwan. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

What does this phase mean?

Phase 4 studies take place after a treatment is approved, to learn more about its long-term benefits and risks in everyday use.

Not every study follows the same path, and some studies (observational studies, device studies) have no phase.

Are results available for this study?

Yes. Results are posted in the registry (first posted 17 Feb 2014). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Best Corrected logMAR Contrast Acuity at Photopic/Mesopic Condition (time frame: average data of post-op 3rd, 6th, 12th week measurements). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Feb 2014.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in